Duong Nguyet M, Nguyen Vu Nhu Q, Le Hai T
Pediatrics Department, Ho Chi Minh City Eye Hospital, Ho Chi Minh City, VNM.
Ophthalmology Department, IVISION Eye Center, Ho Chi Minh City, VNM.
Cureus. 2024 Nov 15;16(11):e73783. doi: 10.7759/cureus.73783. eCollection 2024 Nov.
Purpose The purpose of this study was to compare the accuracy and agreement in nuclear cataract opacity classification between the Smart Eye Camera (SEC) device, and the slit-lamp. Methods A cross-sectional study was conducted with a convenient sampling of 221 eyes from 139 patients with phakic eyes, visiting IVISION Eye Center, Ho Chi Minh City, Vietnam, from November 1, 2023, to April 30, 2024. Two grading systems, Lucio-Buratto (Buratto) and World Health Organization (WHO), were used to compare the effectiveness of the two devices, with statistical analysis using Spearman's correlation coefficient and Cohen's kappa to evaluate the agreement level. Results Results showed no significant difference in Buratto and WHO grading between the two devices (p > 0.05), indicating consistency between the two measurement methods. Spearman's correlation coefficient demonstrated a strong correlation between the results from both devices, with r = 0.797 for Buratto and r = 0.579 for WHO (p < 0.001). The reliability was confirmed by high weighted-kappa values (k = 0.774 for Buratto and k = 0.539 for WHO). Conclusion The SEC's comparable effectiveness to the slit-lamp supports its potential utility in blindness-prevention screening efforts.
目的 本研究旨在比较智能眼相机(SEC)设备与裂隙灯在核性白内障混浊分级方面的准确性和一致性。方法 采用横断面研究,对2023年11月1日至2024年4月30日期间到访越南胡志明市IVISION眼科中心的139例有晶状体眼患者的221只眼睛进行方便抽样。使用两种分级系统,即卢西奥 - 布拉托(Buratto)分级系统和世界卫生组织(WHO)分级系统,来比较这两种设备的有效性,并使用Spearman相关系数和Cohen卡方进行统计分析以评估一致性水平。结果 结果显示两种设备在Buratto分级和WHO分级方面无显著差异(p>0.05),表明两种测量方法具有一致性。Spearman相关系数显示两种设备的结果之间具有强相关性,Buratto分级的r = 0.797,WHO分级的r = 0.579(p<0.001)。通过高加权卡方值(Buratto分级的k = 0.774,WHO分级的k = 0.539)证实了可靠性。结论 SEC与裂隙灯相当的有效性支持了其在预防失明筛查工作中的潜在效用。