Implant effects on polyurethane and silicone cardiac pacing leads in humans: insulation measurements and SEM observations.

作者信息

Barbaro V, Bosi C, Caiazza S, Chistolini P, Ialongo D, Rosa P

出版信息

Biomaterials. 1985 Jan;6(1):28-32. doi: 10.1016/0142-9612(85)90034-1.

Abstract

Recently, polyurethane cardiac pacing leads have been under discussion because of some failures, probably due to modification of surface insulating properties. In order to verify the reliability of polyurethane versus silicone rubber as coating material, the authors, starting from previous clinical research, have carried out a study of electrical insulation related to implant time. At the same time the outer and inner lead surfaces have been submitted to scanning electron microscope (SEM) analysis. The results show, for the polyurethane leads, a significant increase in the leakage or currents in explanted samples, and a significant degradation (cracking) of the outer surfaces.