Roti Elizabeth V, Ni Shuibin, Bayhaqi Yakub, Ostmo Susan R, Burt Spencer S, Woodward Mani K, Jackson John D, Coyner Aaron S, Chiang Michael F, Huang David, Campbell J Peter, Jian Yifan, Young Benjamin K
Casey Eye Institute, Oregon Health and Science University, Portland.
National Eye Institute, National Institutes of Health, Bethesda, Maryland.
JAMA Ophthalmol. 2025 Feb 1;143(2):165-170. doi: 10.1001/jamaophthalmol.2024.5533.
Capturing high-quality images of the entire peripheral retina while minimizing the use of scleral depression could increase the quality of examinations for retinopathy of prematurity (ROP) while reducing neonatal stress.
To evaluate whether an investigational handheld ultra-widefield optical coherence tomography (UWF-OCT) device without scleral depression can be used to document high-quality images of the peripheral retina for use in ROP examinations.
DESIGN, SETTING, AND PARTICIPANTS: This was a prospective, cross-sectional study in the neonatal intensive care unit at a single academic medical center. Study participants were neonates in the neonatal intensive care unit who met standard ROP screening criteria (birth weight ≤1500 g; gestational age ≤30 weeks). Examinations were performed from January 2, 2023, to May 12, 2024. Data analysis was performed in July 2024.
An investigational, contact-based, 800-kHz, 140° (visual angle) UWF-OCT field-of-view handheld device was used during regular ROP examinations.
Images from each examination were captured using the UWF-OCT device, and the zone, stage, and presence of plus disease were diagnosed. If the UWF-OCT system was unable to capture diagnosis, this was recorded.
A total of 507 examinations were performed on 83 neonates (mean [SD] birth weight, 905.1 [362.5] g; mean [SD] gestational age, 189.5 [19.1] days) with this system. The vascular border and disease was able to be visualized in 100% of cases, including to and beyond zone III. In some cases, images were able to clearly demonstrate the ora serrata and even the ciliary body.
The UWF-OCT device was able to be safely used to perform ROP screenings and subsequent evaluations and provided volumetric imaging of the ora serrata and sometimes the ciliary body. The findings suggest that such devices may be associated with reduced technical challenges of capturing the entire peripheral retina in telemedicine programs, may reduce pain and stress for delicate neonates owing to the reduced use of scleral depression, and, with adaptation of the system, could potentially allow for peripheral imaging of the whole retina in older children and adults.
在尽量减少使用巩膜压迫的同时获取整个周边视网膜的高质量图像,可提高早产儿视网膜病变(ROP)检查的质量,同时减轻新生儿的应激反应。
评估一种无需巩膜压迫的手持式超广角光学相干断层扫描(UWF - OCT)研究设备是否可用于记录周边视网膜的高质量图像,以用于ROP检查。
设计、地点和参与者:这是一项在单一学术医疗中心的新生儿重症监护病房进行的前瞻性横断面研究。研究参与者为符合标准ROP筛查标准(出生体重≤1500 g;胎龄≤30周)的新生儿重症监护病房中的新生儿。检查于2023年1月2日至2024年5月12日进行。数据分析于2024年7月进行。
在常规ROP检查期间使用一种研究性的、基于接触的、800 kHz、140°(视角)UWF - OCT视野手持式设备。
每次检查使用UWF - OCT设备采集图像,并诊断病变区域、阶段及有无附加病变。如果UWF - OCT系统无法进行诊断,则记录下来。
使用该系统对83名新生儿(平均[标准差]出生体重905.1[362.5]g;平均[标准差]胎龄189.5[19.1]天)共进行了507次检查。100%的病例能够观察到血管边界和病变,包括Ⅲ区及以外区域。在某些情况下,图像能够清晰显示锯齿缘甚至睫状体。
UWF - OCT设备能够安全地用于进行ROP筛查及后续评估,并提供锯齿缘甚至有时包括睫状体的容积成像。研究结果表明,此类设备可能会减少远程医疗项目中获取整个周边视网膜图像的技术挑战,由于减少了巩膜压迫的使用,可能会减轻脆弱新生儿的疼痛和应激反应,并且随着系统的改进,有可能实现对大龄儿童和成人整个视网膜的周边成像。