Accuracy of the Hounsfield Unit Values Measured by Implant Planning Software.

作者信息

Nagata Koudai, Kouzai Yusuke, Inaba Keitaro, Fujii Manabu, Atsumi Mihoko, Kimoto Katsuhiko, Kuroda Shinji, Kawana Hiromasa

机构信息

Department of Regenerative Implant Dentistry, Kanagawa Dental University, Yokosuka 238-8580, Japan.

Department of Education Planning, Kanagawa Dental University, Yokosuka 238-858, Japan.

出版信息

Dent J (Basel). 2024 Dec 17;12(12):413. doi: 10.3390/dj12120413.

Abstract

: The measurement of Hounsfield units (HU) during implant treatment planning is important. Currently, various manufacturers' implant planning software programs offer HU capabilities; however, their accuracy remains unverified. In this study, we aimed to validate the accuracy of HU values measured by implant planning software programs. : This study used one type of multidetector computed tomography (MDCT), two types of cone-beam computed tomography (CBCT), and four implant planning software packages. Three specimens were prepared for the evaluation of HUs, and the standard values of the HUs were measured. Digital Imaging and Communications in Medicine (DICOM) data obtained from MDCT and CBCT were loaded into four implant planning software packages to measure the HU values. The HU and reference values of the four implant planning software programs obtained from MDCT and CBCT were compared. Additionally, the HU values between each implant planning software program were compared. : The HU values of the three specimens, as measured using the four implant planning software programs utilizing MDCT, did not exhibit a significant difference from the standard values. Conversely, those obtained from CBCT were significantly different. The measured HU values after the MDCT imaging of the specimens were not significantly different between the implant planning software programs; however, they differed after CBCT imaging. : The results of this study indicate that it is not possible to measure HU values using CBCT with implant planning software programs. However, HU values can be measured by any implant planning software using MDCT.

摘要

在种植治疗计划制定过程中,测量亨氏单位(HU)很重要。目前,各制造商的种植计划软件程序都具备测量HU的功能;然而,其准确性尚未得到验证。在本研究中,我们旨在验证种植计划软件程序测量的HU值的准确性。

本研究使用了一种类型的多层螺旋计算机断层扫描(MDCT)、两种类型的锥形束计算机断层扫描(CBCT)以及四个种植计划软件包。制备了三个样本用于评估HU,并测量了HU的标准值。将从MDCT和CBCT获得的医学数字成像和通信(DICOM)数据加载到四个种植计划软件包中以测量HU值。比较了从MDCT和CBCT获得的四个种植计划软件程序的HU值与参考值。此外,还比较了每个种植计划软件程序之间的HU值。

使用四个利用MDCT的种植计划软件程序测量的三个样本的HU值与标准值相比无显著差异。相反,从CBCT获得的HU值则有显著差异。样本经MDCT成像后,种植计划软件程序之间测量的HU值无显著差异;然而,经CBCT成像后则有所不同。

本研究结果表明,使用CBCT和种植计划软件程序无法测量HU值。然而,使用MDCT时,任何种植计划软件都可以测量HU值。