白内障手术中图像引导系统产生的撕囊参考圆直径偏差的相关特征和因素:一项横断面研究。

Characteristics and factors associated with diameter deviation of rhexis reference circle generated by an image-guided system in cataract surgery: a cross-sectional study.

作者信息

Fu Yabo, Huang Yulong, Liu Xiaobao, Kang Chishan, Wu Wenjie, Li Qiong

机构信息

Department of Ophthalmology, Fujian Provincial Hospital, Fuzhou University Affiliated Provincial Hospital, 134 East Street, Fuzhou, 350000, People's Republic of China.

Department of Ophthalmology, Longyan People Hospital of Fujian, Longyan, People's Republic of China.

出版信息

Sci Rep. 2024 Dec 30;14(1):31876. doi: 10.1038/s41598-024-83137-5.

Abstract

More attention should be paid to the deviations between the actual diameter of the rhexis reference circle which projected by the Image-guided systems and its intended size, and assess the influence of ocular biometric parameters on the deviation. In this study, the Callisto eye image-guided system was employed to generate a digital rhexis reference circle (rhexis overlay) set at an intended diameter of 6 mm and a screenshot of the video was taken at the end of the cataract surgery, then to compare the deviation between the observed rhexis overlay diameter (ROD) and the optic diameter (6 mm). The factors influencing diameter deviation were identified with univariate and multivariate linear regression. Subgroup analyses were conducted based on factors identified in the multivariate regression. And we observed that a total of 213 eyes of 174 patients were included in the final analysis. The diameter deviation averaged - 0.24 ± 0.28 mm (P = 0.0001) and was positively correlated with phakic anterior chamber depth (ACD) (P < 0.001) and negatively correlated with white-to-white (WTW) measurement (P < 0.001). Subgroup analyses revealed significantly greater diameter deviation in cases with preoperative ACD < 2.5 mm and WTW ≥ 12.0 mm (all P < 0.05). Following optimization, the mean diameter deviation was 0.00047 ± 0.26 mm (P > 0.05). In conclusion, the Callisto eye image-guided system demonstrates good precision in CCC assistance. However, there remains a slight deviation from the intended size when generating the rhexis reference circle in eyes with shallow anterior chambers and larger corneas. Implementing an optimization process could be beneficial for these eyes.

摘要

应更加关注图像引导系统投射的撕囊参考圆实际直径与其预期大小之间的偏差,并评估眼部生物测量参数对该偏差的影响。在本研究中,采用Callisto眼图像引导系统生成预定直径为6mm的数字撕囊参考圆(撕囊覆盖层),并在白内障手术结束时拍摄视频截图,然后比较观察到的撕囊覆盖层直径(ROD)与视盘直径(6mm)之间的偏差。通过单变量和多变量线性回归确定影响直径偏差的因素。基于多变量回归中确定的因素进行亚组分析。我们观察到,最终分析纳入了174例患者的213只眼。直径偏差平均为-0.24±0.28mm(P=0.0001),与有晶状体眼的前房深度(ACD)呈正相关(P<0.001),与白对白(WTW)测量值呈负相关(P<0.001)。亚组分析显示,术前ACD<2.5mm且WTW≥12.0mm的病例中,直径偏差明显更大(所有P<0.05)。优化后,平均直径偏差为0.00047±0.26mm(P>0.05)。总之,Callisto眼图像引导系统在连续环形撕囊辅助方面显示出良好的精度。然而,在浅前房和大角膜的眼中生成撕囊参考圆时,与预期大小仍存在轻微偏差。实施优化过程可能对这些眼睛有益。