O'Donnell M, Edelstein W A
Med Phys. 1985 Jan-Feb;12(1):20-6. doi: 10.1118/1.595732.
The relative merits of the spin-warp method and the multiple-angle projection-reconstruction method are assessed for nuclear magnetic resonance imaging in the presence of magnetic field inhomogeneities. The results of computer modeling studies demonstrate the superiority of the spin-warp method under these conditions. Because the form of the point spread function for the spin-warp approach is insensitive to magnetic field irregularities, simple correction algorithms have been developed to reduce artifacts associated with bias field inhomogeneities. In addition, the formalism developed for magnetic field inhomogeneities can be easily extended to correct for gradient field nonlinearities.
针对存在磁场不均匀性情况下的核磁共振成像,评估了自旋扭曲法和多角度投影重建法的相对优点。计算机建模研究结果表明在这些条件下自旋扭曲法具有优越性。由于自旋扭曲法的点扩散函数形式对磁场不规则性不敏感,已开发出简单的校正算法以减少与偏置场不均匀性相关的伪影。此外,为磁场不均匀性开发的形式体系可轻松扩展以校正梯度场非线性。