精确的T1和自旋密度核磁共振图像。

Accurate T1 and spin density NMR images.

作者信息

Schneiders N J, Ford J J, Bryan R N

出版信息

Med Phys. 1985 Jan-Feb;12(1):71-6. doi: 10.1118/1.595790.

Abstract

We present a method for producing accurate calculated T1 and spin density nuclear magnetic resonance images. A modified Carr-Purcell-Meiboom-Gill pulse sequence is used to obtain a series of images containing both T1 and T2 dependence. The image series is first analyzed to remove the T2 dependence. The resulting images are then analyzed, pixel by pixel, to generate an image containing T1 values and an image containing values proportional to spin density (SD). Tests performed on two phantoms containing solutions of various known T1's and H2O/D2O concentrations indicate that the T1 image values are accurate to better than 11% and the relative SD values agree to within one standard deviation.

摘要

我们提出了一种用于生成精确计算的T1和自旋密度核磁共振图像的方法。使用一种改进的Carr-Purcell-Meiboom-Gill脉冲序列来获取一系列包含T1和T2依赖性的图像。首先对该图像序列进行分析以消除T2依赖性。然后逐像素地分析所得图像,以生成包含T1值的图像和包含与自旋密度(SD)成比例的值的图像。对含有各种已知T1值和H2O/D2O浓度溶液的两个体模进行的测试表明,T1图像值的准确度优于11%,相对SD值在一个标准差范围内相符。