Evaluation of a Point-of-Care-Viscoelastic Coagulation Device in Hispaniolan Amazon Parrots ().

作者信息

Russell Laken, Tully Thomas N, Smith M Ryan, Mitchell Mark A

机构信息

Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803, USA,

Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803, USA.

出版信息

J Avian Med Surg. 2025 Jan;38(4):188-196. doi: 10.1647/AVIANMS-D-23-00050.

Abstract

Viscoelastic testing evaluates the formation and lysis of a clot over time, allowing more complete assessment of patient hemostasis in real time, whereas traditional tests, such as prothrombin time and partial thromboplastin time, only measure coagulation factor function. Patient-side viscoelastic coagulation monitors are easy to use, portable, and provide faster turnaround time than commercial laboratories. Viscoelastic testing requires only 0.2 mL of blood and is useful in diagnosing and treating hemostatic disorders. Currently, there is no standardized coagulation testing method across bird species. In this cross-sectional study, a viscoelastic coagulation device, the Entegrion Viscoelastic Coagulation Monitor-Vet (VCM-Vet), was evaluated. Blood samples were obtained from 26 Hispaniolan Amazon parrots (HAPs) () under manual restraint. Results were recorded on the device as graphical output with quantitative viscoelastic measurements. Results were reported using standard rotational thromboelastometry terminology, including clotting time, clot formation time, alpha angle, maximum clot firmness, clot firmness amplitude at 10 and 20 minutes after clot formation, and clot lysis at 30 and 45 minutes. The median clotting time was 463 seconds (reference interval: 56-1635 seconds), the mean clot formation time was 704.7 seconds (reference interval: 172-1697 seconds), the mean alpha angle was 27.3 (reference interval: 7-60), and the mean maximum clot firmness was 15.4 (reference interval: 7-25). Statistical analysis found that all parameters were normally distributed aside from clotting time in seconds. There was no appreciable breakdown of the clot during the 60-minute device runtime, and there was no significant difference in any parameter based on sex. The VCM-Vet produced clotting times for this population of HAPs and enabled the creation of reference intervals. Based on our findings, the VCM-Vet can be used to assess clot potential in HAPs and possibly other avian species.

摘要

黏弹性检测可评估随时间形成和溶解的血凝块,从而能更全面地实时评估患者的止血情况,而传统检测方法,如凝血酶原时间和活化部分凝血活酶时间,仅测量凝血因子功能。患者端黏弹性凝血监测仪使用方便、便于携带,且比商业实验室能更快给出检测结果。黏弹性检测仅需0.2毫升血液,对止血障碍的诊断和治疗很有用。目前,鸟类中尚无标准化的凝血检测方法。在这项横断面研究中,对一种黏弹性凝血设备——恩特格瑞恩黏弹性凝血监测仪-兽医版(VCM-Vet)进行了评估。在人工约束下从26只伊斯帕尼奥拉亚马逊鹦鹉(HAPs)采集血样。结果以图形输出形式记录在该设备上,并进行定量黏弹性测量。结果使用标准旋转血栓弹力图术语报告,包括凝血时间、凝血形成时间、α角、最大血凝块硬度、凝血形成后10分钟和20分钟时的血凝块硬度振幅,以及30分钟和45分钟时的血凝块溶解情况。中位凝血时间为463秒(参考区间:56 - 1635秒),平均凝血形成时间为704.7秒(参考区间:172 - 1697秒),平均α角为27.3(参考区间:7 - 60),平均最大血凝块硬度为15.4(参考区间:7 - 25)。统计分析发现,除了以秒为单位的凝血时间外,所有参数均呈正态分布。在设备运行的60分钟内,血凝块没有明显分解,且基于性别在任何参数上均无显著差异。VCM-Vet为这群HAPs生成了凝血时间,并能够创建参考区间。基于我们的研究结果,VCM-Vet可用于评估HAPs以及可能其他鸟类的凝血潜力。

相似文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验