保罗青光眼植入物和艾哈迈德ClearPath 250在管内有无聚丙烯线情况下的耐压性体外分析。

In Vitro Analysis of Pressure Resistance in the Paul Glaucoma Implant and Ahmed ClearPath 250 With and Without Polypropylene Thread Inside the Tube.

作者信息

Masdipa Andi, Kaidzu Sachiko, Tanito Masaki

机构信息

Department of Ophthalmology, Shimane University Faculty of Medicine, Izumo, Japan.

出版信息

Transl Vis Sci Technol. 2025 Jan 2;14(1):2. doi: 10.1167/tvst.14.1.2.

Abstract

PURPOSE

Pressure resistance characteristics of the Paul glaucoma implant (PGI) and Ahmed ClearPath 250 (ACP), with and without the insertion of polypropylene thread in their tubes, were evaluated.

METHODS

The in vitro flow pressure was evaluated at varying flow rates, both with and without threads (6-0 for PGI and 4-0 or 3-0 for ACP). Cross-sectional areas of the tube lumen and thread were measured to calculate pressure resistance using the Hagen-Poiseuille equation.

RESULTS

For the PGI without the thread, the pressure remained relatively low and constant across all flow rates. In contrast, with the insertion of a 6-0 thread, there was a significant increase in pressure resistance, with the pressure increasing from 7.5 mm Hg at 1 µL/min to 43.8 mm Hg at 5 µL/min. For the ACP, the pressure resistance remained relatively constant across all flow rates without a thread and with either 4-0 or 3-0 threads. However, the pressure was higher with 3-0 threads compared with a 4-0 thread. The actual measured pressures agreed well with theoretical values in the no-thread conditions, but were consistently higher than the theoretical values in the threaded conditions.

CONCLUSIONS

Inserting polypropylene threads into the tubes of nonvalved glaucoma drainage devices significantly affects pressure resistance with various degree.

TRANSLATIONAL RELEVANCE

PGI with a 6-0 polypropylene thread may not require external tube ligation to prevent hypotony, whereas ACP with a 4-0 thread likely requires additional ligation. Using a 3-0 thread in ACP may enhance pressure resistance sufficiently to avoid tube ligation, but this requires careful clinical consideration.

摘要

目的

评估保罗青光眼植入物(PGI)和艾哈迈德ClearPath 250(ACP)在其引流管插入和未插入聚丙烯线情况下的耐压特性。

方法

在有和没有线(PGI用6-0线,ACP用4-0或3-0线)的情况下,于不同流速下评估体外流动压力。测量引流管腔和线的横截面积,使用哈根-泊肃叶方程计算耐压性。

结果

对于未插入线的PGI,在所有流速下压力都保持相对较低且恒定。相比之下,插入6-0线后,耐压性显著增加,压力从1微升/分钟时的7.5毫米汞柱增加到5微升/分钟时的43.8毫米汞柱。对于ACP,在未插入线以及插入4-0或3-0线的所有流速下,耐压性都保持相对恒定。然而,使用3-0线时的压力高于使用4-0线时的压力。在无线情况下实际测量的压力与理论值吻合良好,但在有线情况下始终高于理论值。

结论

在无阀门的青光眼引流装置的引流管中插入聚丙烯线会不同程度地显著影响耐压性。

转化相关性

带有6-0聚丙烯线的PGI可能不需要外部引流管结扎来预防低眼压,而带有4-0线的ACP可能需要额外结扎。在ACP中使用3-0线可能会充分增强耐压性以避免引流管结扎,但这需要仔细的临床考量。