La Rosa Giusy Rita Maria, Lombardo Pietro, Generali Luigi, Pedullà Eugenio
Department of General Surgery and Medical-Surgical Specialties, University of Catania, 95123, Catania, Italy.
Department of Surgery, Medicine, Dentistry and Morphological Sciences With Transplant Surgery, Oncology and Regenerative Medicine Relevance (CHIMOMO), University of Modena and Reggio Emilia, 41124, Modena, Italy.
Odontology. 2025 Jan 13. doi: 10.1007/s10266-024-01051-8.
The aim of this study was to assess the cyclic fatigue resistance of a single-file system (i.e., Hyflex EDM OneFile), during continuous rotation and reflex dynamic motion with and without irrigation. Cyclic fatigue tests were conducted on 48 new Hyflex EDM One files using two different motions, with and without irrigation. The files were randomly assigned to four groups (n = 12) based on the tested motion and irrigation conditions: continuous rotation and the novel kinematics ReFlex Dynamic motion, with and without irrigation. The dynamic fatigue resistance was measured as the number of cycles to fracture (NCF) in an artificial zirconium canal with a 60° angle and a 5 mm radius of curvature at an intracanal temperature. The results were analyzed using the non-parametric Mann-Whitney test, with a significance level established at 5% (p < 0.05). The fractured instruments were examined with a scanning electron microscope. The instruments activated in continuous rotation, irrespective of irrigation conditions, demonstrated significantly higher NCF compared to those in ReFlex Dynamic (p < 0.05). Irrigation significantly impacted the cyclic fatigue resistance of Hyflex EDM in ReFlex Dynamic motion (p < 0.05), while it did not influence the resistance of files tested under continuous rotation (p > 0.05). Within the limitations of this in vitro study, HyFlex EDM One files tested under continuous rotation showed greater resistance to cyclic fatigue compared to those tested with ReFlex Dynamic motion, regardless of irrigation. Continuous hypochlorite irrigation enhanced the cyclic fatigue resistance only of the files tested in ReFlex Dynamic motion.
本研究的目的是评估一种单根管系统(即Hyflex EDM OneFile)在连续旋转以及有无冲洗条件下的往复动态运动中的抗循环疲劳性能。使用两种不同的运动方式,对48支全新的Hyflex EDM One锉进行了循环疲劳测试,测试过程中有冲洗和无冲洗两种情况。根据测试的运动方式和冲洗条件,将锉随机分为四组(n = 12):连续旋转以及新型运动学的ReFlex动态运动,每组又分有冲洗和无冲洗两种情况。动态抗疲劳性能通过在人工锆管中,在根管内温度条件下,以60°角和5毫米曲率半径折断时的循环次数(NCF)来衡量。结果采用非参数曼-惠特尼检验进行分析,显著性水平设定为5%(p < 0.05)。对折断的器械用扫描电子显微镜进行检查。无论冲洗条件如何,连续旋转激活的器械与ReFlex动态运动的器械相比,其NCF显著更高(p < 0.05)。冲洗对ReFlex动态运动中Hyflex EDM的抗循环疲劳性能有显著影响(p < 0.05),而对连续旋转测试的器械的抗疲劳性能没有影响(p > 0.05)。在本体外研究的局限性范围内,无论有无冲洗,连续旋转测试的HyFlex EDM One锉比ReFlex动态运动测试的锉表现出更强的抗循环疲劳能力。连续次氯酸钠冲洗仅增强了ReFlex动态运动测试的锉的抗循环疲劳性能。