新型CT探测器与过滤技术的评估:第1部分 - X射线束特性及辐射剂量测定

Assessment of a New CT Detector and Filtration Technology: Part 1 - X-ray Beam Characterization and Radiation Dosimetry.

作者信息

Barreto Izabella, Correa Nathalie, Moser Patricia, Tuna Ibrahim, Massini Tara, Rill Lynn, Arreola Manuel

机构信息

Department of Radiology, College of Medicine, University of Florida, Gainesville, FL.

出版信息

J Comput Assist Tomogr. 2025;49(4):625-630. doi: 10.1097/RCT.0000000000001710. Epub 2025 Jan 28.

Abstract

PURPOSE

This study evaluated beam quality and radiation dosimetry of a CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO). PVO features miniaturized electronics, a detector cut with microblade technology, and increased filtration in order to increase x-ray detection and reduce image noise.

METHODS

We assessed the performance of two similar 320-detector CT scanners: one equipped with PVO and one without. Beam quality was measured by determining the half-value layer (HVL) and effective energy ( Eeff ) for both scanners using all tube voltages (80 kV, 100 kV, 120 kV, 135 kV) and bowtie filters (small, medium, large) available. Energy correction factors were identified for optically stimulated luminescent dosimeters (OSLDS) compared to a calibrated ionization chamber. Surface and internal doses were measured for an anthropomorphic CT angiography head phantom and a cadaver head scanned with CTDI vol matched as close as possible at the conventional (55.1 mGy) and PVO (55.4 mGy) CT scanners.

RESULTS

For all scan settings, the PVO scanner showed significantly higher HVL (range, 4.33-11.02 mm Al) and effective energy (range, 39.4-68.0 keV) values compared to the conventional scanner (HVL, 4.19-8.25 mm Al; effective energy, 38.4-55.2 keV). For equivalent CTDI vol values, the energy-corrected surface skin and lens doses were on average 6.7% lower with the PVO scanner than the conventional scanner ( P  < 0.01).

CONCLUSIONS

PVO technology yielded higher HVL and effective energies and, for the same CTDI vol , resulted in lower surface organ doses, indicating a potential for reduced patient radiation exposure in clinical settings.

摘要

目的

本研究评估了一台配备名为PureVision Optics(PVO)的新型探测器和过滤技术的CT扫描仪的射束质量和辐射剂量学。PVO具有小型化电子设备、采用微刀片技术切割的探测器以及增加的过滤功能,以提高X射线检测并降低图像噪声。

方法

我们评估了两台类似的320探测器CT扫描仪的性能:一台配备PVO,另一台未配备。通过使用所有可用的管电压(80 kV、100 kV、120 kV、135 kV)和蝴蝶结滤波器(小、中、大)确定两台扫描仪的半值层(HVL)和有效能量(Eeff)来测量射束质量。与校准电离室相比,确定了光激励发光剂量计(OSLDS)的能量校正因子。在传统(55.1 mGy)和PVO(55.4 mGy)CT扫描仪上,对一个人体头部CT血管造影模型和一个尸体头部进行扫描,测量表面剂量和内部剂量,并尽可能使CTDI vol匹配。

结果

对于所有扫描设置,与传统扫描仪(HVL,4.19 - 8.25 mm Al;有效能量,38.4 - 55.2 keV)相比,PVO扫描仪显示出显著更高的HVL值(范围,4.33 - 11.02 mm Al)和有效能量值(范围,39.4 - 68.0 keV)。对于等效的CTDI vol值,PVO扫描仪的能量校正表面皮肤和晶状体剂量平均比传统扫描仪低6.7%(P < 0.01)。

结论

PVO技术产生了更高的HVL和有效能量,并且在相同的CTDI vol下,导致更低的表面器官剂量,表明在临床环境中有可能降低患者的辐射暴露。