Calvo R, Carlos R, Erill S
Int J Clin Pharmacol Ther Toxicol. 1985 Feb;23(2):76-8.
Phenylbutazone and clofibric acid, two drugs strongly bound to human albumin, produce low readings of albumin content in serum when the bromocresol green immediate reaction is used. This abnormality is observed at drug concentrations within the range obtained during therapeutic use, and tends to be more marked in diluted samples of serum. Abnormally low values of albumin content are also obtained when the bromocresol green method is used in uremic sera, and the disparity seems related to the degree of carbamylation of these samples. The reported interferences are great enough in some cases as to suggest that the use of the immediate reaction between bromocresol green and serum should not be considered a valid measure of albumin content when these factors cannot be totally excluded.
保泰松和氯贝酸这两种与人体白蛋白紧密结合的药物,在使用溴甲酚绿直接反应法时,会使血清中白蛋白含量读数偏低。在治疗用药期间所达到的药物浓度范围内可观察到这种异常情况,并且在稀释的血清样本中往往更为明显。当在尿毒症血清中使用溴甲酚绿法时,也会得到异常低的白蛋白含量值,这种差异似乎与这些样本的氨甲酰化程度有关。在某些情况下,所报道的干扰非常大,以至于当这些因素不能完全排除时,不应将溴甲酚绿与血清之间的直接反应视为测定白蛋白含量的有效方法。