Technical note: Evaluation of a transportable linear ion trap mass spectrometer for seized drug screening operated under controlled laboratory conditions.

作者信息

Couch Alleigh N, Chang Christina, Davidson J Tyler

机构信息

Department of Forensic Science, Sam Houston State University, Huntsville, TX, USA.

Department of Forensic Science, Sam Houston State University, Huntsville, TX, USA.

出版信息

Forensic Sci Int. 2025 Feb;367:112381. doi: 10.1016/j.forsciint.2025.112381. Epub 2025 Jan 25.

Abstract

Increasing seized drug casework submissions and challenges with rapidly triaging seized drug evidence have led to interest in the potential use of instruments for on-site or mobile laboratory drug screening. However, before this type of instrumentation can be implemented, it is imperative to evaluate the capabilities and limitations under controlled laboratory conditions. This study evaluates the BaySpec Continuity™ transportable linear ion trap (LIT) mass spectrometer coupled to a thermal desorption-atmospheric pressure chemical ionization (TD-APCI) source for seized drug screening while operated under controlled laboratory conditions. The target compounds chosen for this study were selected based on controlled substances and cutting agents commonly observed in seized drug casework, including cocaine, methamphetamine, heroin, fentanyl, caffeine, acetaminophen, pseudoephedrine, phenacetin, phentermine, and levamisole. A library-based alarm system screening method was developed and validated for rapid compound identification in seized drug mixtures. Selectivity, repeatability, and reproducibility were evaluated using five simulated seized drug mixtures with known compositions, and the limit of detection was evaluated for each compound. Blind and authentic destroyed casework samples were analyzed to determine the efficacy of this method when applied to real-world samples and a 100% correct identification rate of the controlled substance present was achieved. Additionally, the ability to differentiate isobaric species using tandem mass spectrometry (MS/MS) analysis was demonstrated for methamphetamine and phentermine. The validated library-based alarm screening method shows promise for rapid seized drug screening under controlled laboratory conditions, which supports potential usage in a mobile laboratory setting.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验