机器人辅助与人工全膝关节置换术中与噪声相关的损伤

Noise-Related Injury During Robotic versus Manual Total Knee Arthroplasty.

作者信息

Pagan Dianne N, Cruz-Ossa Natalia, Cade William, D'Apuzzo Michele R, Carvajal Jaime A, Hernandez Victor H

机构信息

Department of Orthopedic Surgery, University of Miami Miller School of Medicine, Miami, Florida.

出版信息

J Arthroplasty. 2025 Aug;40(8):2042-2047. doi: 10.1016/j.arth.2025.02.022. Epub 2025 Feb 14.

Abstract

BACKGROUND

Occupational noise exposure during robotic total knee arthroplasty (TKA) poses a risk to orthopaedic surgeons that has not been quantified in the literature. According to National Institute for Occupational Safety and Health (NIOSH) guidelines, noise exposure levels at or above 85 decibels as an 8-hour time-weighted average (TWA) are hazardous. The objectives of this study were to compare noise metrics between manual and robotic TKAs and determine if the TWA of each group exceeds the NIOSH guidelines required for workplace noise safety.

METHODS

Intraoperative audio was recorded during manual and two different robotic TKAs. The DecibelX application was used to record the duration of exposure, average and maximum decibel level, TWA, noise dose percentage of maximum allowable daily noise, and noise dose percentage projected forward over eight hours. Dwass-Steel-Critchlow-Fligner and Kruskal-Wallis tests were performed. Statistical significance was set at P < 0.05.

RESULTS

There were 66 recordings analyzed. The mean duration of exposure and mean maximum decibels were comparable between the groups. The mean average decibel level was 92.7 dB for manual TKA, 97.6 dB for robotic TKA system A, and 92.7 dB for robotic TKA system B (P < 0.01). The mean noise dose for manual TKA was 114.1, 235.7 for robotic TKA system A and 83.6% for robotic TKA system B (P < 0.01). The mean projected noise dose for manual TKA was 818.9, 1,936.3 for robotic TKA system A and 667.9% for robotic TKA system B (P < 0.01). The mean TWA was 83.8 dB for manual TKA, 88.3 dB for robotic TKA system A, and 82.8 dB for robotic TKA system B (P < 0.01).

CONCLUSIONS

With a mean TWA of 88.3 dB, robotic TKA system A procedure is above the limit of 85 dB set by NIOSH occupational noise exposure guidelines and is in the range of noise levels that are known to cause noise-induced hearing loss.

摘要

背景

机器人全膝关节置换术(TKA)过程中的职业噪声暴露对骨科医生构成风险,而这一风险在文献中尚未得到量化。根据美国国家职业安全与健康研究所(NIOSH)的指南,8小时时间加权平均(TWA)噪声暴露水平达到或超过85分贝是有害的。本研究的目的是比较手动和机器人TKA的噪声指标,并确定每组的TWA是否超过NIOSH工作场所噪声安全指南的要求。

方法

在手动和两种不同的机器人TKA手术过程中记录术中音频。使用DecibelX应用程序记录暴露持续时间、平均和最大分贝水平、TWA、最大允许每日噪声的噪声剂量百分比以及八小时内预测的噪声剂量百分比。进行了Dwass-Steel-Critchlow-Fligner和Kruskal-Wallis检验。设定统计学显著性为P < 0.05。

结果

共分析了66份记录。各组之间的平均暴露持续时间和平均最大分贝数相当。手动TKA的平均分贝水平为92.7 dB,机器人TKA系统A为97.6 dB,机器人TKA系统B为92.7 dB(P < 0.01)。手动TKA的平均噪声剂量为114.1,机器人TKA系统A为235.7,机器人TKA系统B为83.6%(P < 0.01)。手动TKA的平均预测噪声剂量为818.9,机器人TKA系统A为1,936.3,机器人TKA系统B为667.9%(P < 0.01)。手动TKA的平均TWA为83.8 dB,机器人TKA系统A为88.3 dB,机器人TKA系统B为82.8 dB(P < 0.01)。

结论

机器人TKA系统A手术的平均TWA为88.3 dB,高于NIOSH职业噪声暴露指南设定的85 dB限值,处于已知会导致噪声性听力损失的噪声水平范围内。