Zhu DaTian, Wang LuMeng, Qu ZhanXin, Wang Pu, Yuan YongYi, Yang Lan, Yao XinYu, Zheng XiaoBo, Lian HengLi, Wang Chong, Wang JunJie, Ye YuFeng, Elsheikh Ahmed, Bao FangJun, Li JunHua
National Clinical Research Center for Ocular Diseases, Eye Hospital, WenZhou Medical University, Wenzhou, 325027, China.
Eye Center, Xinjiang 474 Hospital, Beijing Road 754, Urimqi, 830000, China.
BMC Ophthalmol. 2025 Feb 26;25(1):90. doi: 10.1186/s12886-025-03913-3.
To assess the diurnal changes in corneal biomechanics in healthy young adults.
At Wenzhou Medical University Eye Hospital in China, a prospective case series. Each healthy subject had six Corvis ST examinations, from 10:00 on a first day to 6:00 on the following day, at 4-hour intervals. The data collected included the biomechanically corrected IOP (bIOP), the central corneal thickness (CCT) and six dynamic corneal response (DCR) parameters, namely deformation amplitude (DA), the ratio between DA values at the apex and 2 mm from the apex (DAR2), integrated inverse radius (IIR), the stiffness parameter at first applanation (SP-A1), Corvis biomechanical index (CBI) and the updated stress-strain index (SSIv2). A multilevel model was used to assess the changes in DCR parameters of patients over time during the examination period.
Sixty-one subjects were included and intra-operator reproducibility was good for most DCR parameters at the 6 time points. DA, DAR2, IIR and CBI showed small fluctuations during the 24-hour cycle with mean values of 0.1, 0.16, 0.34, and 0.04 units, respectively, while SP-A1 showed higher variations averaging 20.57 units and SSIv2 had the least fluctuation within 0.10 units during 24-hour cycle. After correcting for the effects of CCT and bIOP, the diurnal fluctuations in IIR, CBI and SSIv2 were not statistically significant, unlike the fluctuations in DA, DAR2 and SP-A1, which remained significant.
Compared with the DA, DAR2, IIR and CBI, SP-A1 was more affected while SSIv2 was the less affected by the diurnal variations and associated corneal edema.
评估健康年轻成年人角膜生物力学的昼夜变化。
在中国温州医科大学眼视光医院进行一项前瞻性病例系列研究。每位健康受试者在第一天10:00至次日6:00期间,每隔4小时进行6次Corvis ST检查。收集的数据包括生物力学校正眼压(bIOP)、中央角膜厚度(CCT)以及六个动态角膜反应(DCR)参数,即变形幅度(DA)、角膜顶点与距顶点2毫米处DA值的比值(DAR2)、积分逆半径(IIR)、首次压平处的刚度参数(SP-A1)、Corvis生物力学指数(CBI)和更新的应力应变指数(SSIv2)。使用多水平模型评估检查期间患者DCR参数随时间的变化。
纳入61名受试者,多数DCR参数在6个时间点的操作者内重复性良好。DA、DAR2、IIR和CBI在24小时周期内波动较小,平均值分别为0.1、0.16、0.34和0.04单位,而SP-A1变化较大,平均值为20.57单位,SSIv2在24小时周期内波动最小,在0.10单位以内。校正CCT和bIOP的影响后,IIR、CBI和SSIv2的昼夜波动无统计学意义,与DA、DAR2和SP-A1的波动不同,后者仍具有显著性。
与DA、DAR2、IIR和CBI相比,SP-A1受昼夜变化和相关角膜水肿的影响更大,而SSIv2受影响较小。