Wahner H W, Dunn W L, Mazess R B, Towsley M, Lindsay R, Markhard L, Dempster D
Radiology. 1985 Jul;156(1):203-6. doi: 10.1148/radiology.156.1.4001406.
Dual-photon absorptiometry with gadolinium 153 was used to measure the mineral content of lumbar vertebrae in cadavers, excised vertebrae with marrow, and dry, marrow-free vertebrae. The error introduced by the surrounding soft tissue of cadavers was 3%, and the error in determining mineral mass or density in excised vertebrae was about 5%. The correlation coefficient between the results of Gd-153 and corrected iodine 125 (single-photon) absorptiometry on 24 femoral necks was 0.99, and the predictive error was 3.7%. Dual-photon absorptiometry accurately indicates bone mass and bone density and is only slightly affected by either surrounding tissue or fat changes in bone marrow.
使用钆153双能X线吸收法测量尸体腰椎、带骨髓的切除椎体以及干燥无骨髓椎体的矿物质含量。尸体周围软组织引入的误差为3%,在确定切除椎体的矿物质质量或密度时的误差约为5%。钆153与校正后的碘125(单光子)吸收法对24个股骨颈测量结果的相关系数为0.99,预测误差为3.7%。双能X线吸收法能准确显示骨量和骨密度,且仅受周围组织或骨髓脂肪变化的轻微影响。