Pacifici R, Rupich R, Vered I, Fischer K C, Griffin M, Susman N, Avioli L V
Department of Medicine, Jewish Hospital, Washington University Medical Center, St. Louis, Missouri 63110.
Calcif Tissue Int. 1988 Sep;43(3):189-91. doi: 10.1007/BF02571319.
Dual energy radiography (DER) has been recently proposed as a new technique for the non invasive measurement of vertebral mineral density. This preliminary study was undertaken to assess the short-term precision of DER and to compare DER to dual photon absorptiometry (DPA) measurements. Three DER measurements were obtained in 19 healthy volunteers (mean age 34.7 +/- 9.3 years) over a 3 week period. In addition, both DER and DPA measurements were obtained in 52 women (mean age 54.3 +/- 12.5 years). The coefficient of variation for the DER measurements was 1.00%. A highly significant correlation (r = 0.94; p less than 0.0001) was found between DPA and DER. Using this curve a conversion factor of -6.8% was calculated to convert DER to DPA values. Both DER and DPA were significantly and similarly correlated with age (r = -0.43, p less than 0.005; r = -0.38, p less than 0.05, respectively). The rates of bone loss, 0.66% per year for DER and 0.57% per year for DPA, were also similar with the two techniques. From these data, it is concluded that DER is a precise new method for vertebral bone mass measurements.
双能X线摄影术(DER)最近被提议作为一种测量椎体骨密度的无创新技术。本初步研究旨在评估DER的短期精度,并将DER与双光子吸收测定法(DPA)测量结果进行比较。在3周内对19名健康志愿者(平均年龄34.7±9.3岁)进行了3次DER测量。此外,还对52名女性(平均年龄54.3±12.5岁)进行了DER和DPA测量。DER测量的变异系数为1.00%。发现DPA和DER之间存在高度显著的相关性(r = 0.94;p < 0.0001)。利用该曲线计算出将DER转换为DPA值的转换因子为-6.8%。DER和DPA均与年龄显著相关且相似(分别为r = -0.43,p < 0.005;r = -0.38,p < 0.05)。两种技术的骨质流失率也相似,DER为每年0.66%,DPA为每年0.57%。从这些数据得出结论,DER是一种精确的测量椎体骨量的新方法。