比较头颈部定量磁化率成像的重复性指标。

Comparing repeatability metrics for quantitative susceptibility mapping in the head and neck.

作者信息

Cherukara Matthew T, Shmueli Karin

机构信息

Department of Medical Physics and Biomedical Engineering, University College London, London, UK.

出版信息

MAGMA. 2025 Mar 1. doi: 10.1007/s10334-025-01229-3.

Abstract

OBJECTIVE

Quantitative susceptibility mapping (QSM) is a technique that has been demonstrated to be highly repeatable in the brain. As QSM is applied to other parts of the body, it is necessary to investigate metrics for quantifying repeatability, to enable optimization of repeatable QSM reconstruction pipelines beyond the brain.

MATERIALS AND METHODS

MRI data were acquired in the head and neck (HN) region in ten healthy volunteers, who underwent six acquisitions across two sessions. QSMs were reconstructed using six representative state-of-the-art techniques. Repeatability of the susceptibility values was compared using voxel-wise metrics (normalized root mean squared error and XSIM) and ROI-based metrics (within-subject and between-subject standard deviation, coefficient of variation (CV), intraclass correlation coefficient (ICC)).

RESULTS

Both within-subject and between-subject variations were smaller than the variation between QSM dipole inversion methods, in most ROIs. autoNDI produced the most repeatable susceptibility values, with ICC > 0.75 in three of six HN ROIs with an average ICC of 0.66 across all ROIs. Joint consideration of standard deviation and ICC offered the best metric of repeatability for comparisons between QSM methods, given typical distributions of positive and negative QSM values.

DISCUSSION

Repeatability of QSM in the HN region is highly dependent on the dipole inversion method chosen, but the most repeatable methods (autoNDI, QSMnet, TFI) are only moderately repeatable in most HN ROIs.

摘要

目的

定量磁化率成像(QSM)是一种已被证明在脑部具有高度可重复性的技术。随着QSM应用于身体的其他部位,有必要研究用于量化可重复性的指标,以便在脑部以外优化可重复的QSM重建流程。

材料与方法

在10名健康志愿者的头颈部(HN)区域采集MRI数据,他们在两个阶段共进行了6次采集。使用6种具有代表性的先进技术重建QSM。使用体素级指标(归一化均方根误差和XSIM)和基于感兴趣区域(ROI)的指标(受试者内和受试者间标准差、变异系数(CV)、组内相关系数(ICC))比较磁化率值的可重复性。

结果

在大多数ROI中,受试者内和受试者间的变异均小于QSM偶极反转方法之间的变异。autoNDI产生的磁化率值最具可重复性,在6个HN ROI中的3个中ICC>0.75,所有ROI的平均ICC为0.66。考虑到QSM正负值的典型分布,标准差和ICC的联合考量为比较QSM方法提供了最佳的可重复性指标。

讨论

QSM在HN区域的可重复性高度依赖于所选择的偶极反转方法,但最具可重复性的方法(autoNDI、QSMnet、TFI)在大多数HN ROI中仅具有中等可重复性。