Suppr超能文献

股骨颈骨折愈合后取出股骨颈系统对生物力学稳定性及螺钉松动风险的影响。

Impact of femoral neck system removal after femoral neck fracture healing on biomechanical stability and screw stripping risk.

作者信息

Lee Se-Won, Pak Jeongah, Lim Dohyung

机构信息

Department of Orthopedic Surgery, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.

Department of Mechanical Engineering, Sejong University, 98 Gunja-Dong, Gwangjin-Gu, Seoul, 147- 747 Republic of Korea.

出版信息

Biomed Eng Lett. 2024 Dec 27;15(2):349-355. doi: 10.1007/s13534-024-00452-6. eCollection 2025 Mar.

Abstract

This study aims to determine whether the removal of the femoral neck system (FNS) after bony union affects the biomechanical stability of the femur. Considering the technical challenges and potential complications, including screw stripping reported in recent studies, the study explores whether its removal impacts stress distribution within the femur and increases the risk of complications, such as screw stripping. The femurs were grouped into Intact, Group U (healed fractures with FNS in place), and Group R (healed fractures with FNS removed). Subgroup analysis was performed using Pauwels' classification for fractures at 30, 50, and 70 degrees. Finite element analysis (FEA) was used to model and evaluate the biomechanical behavior. Material properties for the models were derived from the literature. No significant difference in biomechanical stability was observed between Group U and Group R across the fracture angles tested, indicating that removal of FNS does not compromise the structural integrity of the femur. However, the risk of screw stripping during removal requires consideration. Removing the femoral neck system (FNS) after fracture healing preserves the femur's biomechanical stability, regardless of fracture angle. However, increased stress at the distal locking screw suggests caution to avoid complications such as screw stripping.

摘要

本研究旨在确定骨愈合后股骨颈系统(FNS)的移除是否会影响股骨的生物力学稳定性。考虑到技术挑战和潜在并发症,包括近期研究中报道的螺钉松动,本研究探讨其移除是否会影响股骨内的应力分布并增加并发症风险,如螺钉松动。股骨被分为完整组、U组(骨折愈合且FNS在位)和R组(骨折愈合且FNS已移除)。使用 Pauwels 分类法对 30度、50度和 70度骨折进行亚组分析。采用有限元分析(FEA)对生物力学行为进行建模和评估。模型的材料属性来自文献。在测试的骨折角度范围内,U组和R组之间未观察到生物力学稳定性的显著差异,表明移除FNS不会损害股骨的结构完整性。然而,移除过程中螺钉松动的风险需要考虑。骨折愈合后移除股骨颈系统(FNS)可保持股骨的生物力学稳定性,无论骨折角度如何。然而,远端锁定螺钉处应力增加提示应谨慎操作以避免诸如螺钉松动等并发症。

相似文献

1
Impact of femoral neck system removal after femoral neck fracture healing on biomechanical stability and screw stripping risk.
Biomed Eng Lett. 2024 Dec 27;15(2):349-355. doi: 10.1007/s13534-024-00452-6. eCollection 2025 Mar.
4
Femoral neck fractures after removal of hardware in healed trochanteric fractures.
Injury. 2017 Dec;48(12):2619-2624. doi: 10.1016/j.injury.2017.11.031.
5
[Analysis of effectiveness of Holosight robot navigation-assisted percutaneous cannulated screw fixation in treatment of femoral neck fractures].
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2025 Jun 15;39(6):673-679. doi: 10.7507/1002-1892.202502067.
6
Are There Differences in Performance Among Femoral Stem Brands Utilized in Cementless Hemiarthroplasty for Treatment of Geriatric Femoral Neck Fractures?
Clin Orthop Relat Res. 2025 Feb 1;483(2):253-264. doi: 10.1097/CORR.0000000000003222. Epub 2024 Aug 15.
7
Cephalomedullary nails versus extramedullary implants for extracapsular hip fractures in older adults.
Cochrane Database Syst Rev. 2022 Jan 26;1(1):CD000093. doi: 10.1002/14651858.CD000093.pub6.

本文引用的文献

2
Fixation of femoral neck fracture with femoral neck system: a retrospective cohort study of 43 patients.
BMC Musculoskelet Disord. 2024 Jan 2;25(1):8. doi: 10.1186/s12891-023-07113-2.
5
Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis.
Eur J Trauma Emerg Surg. 2023 Aug;49(4):1845-1853. doi: 10.1007/s00068-023-02253-5. Epub 2023 Mar 25.
6
Effect of screw tunnels on proximal femur strength after screw removal: A finite element analysis.
Orthop Traumatol Surg Res. 2022 Dec;108(8):103408. doi: 10.1016/j.otsr.2022.103408. Epub 2022 Sep 16.
8
Finite element analysis of femoral neck system in the treatment of Pauwels type III femoral neck fracture.
Medicine (Baltimore). 2022 Jul 15;101(28):e29450. doi: 10.1097/MD.0000000000029450.
9
Fatigue and biological properties of Ti-6Al-4V ELI cellular structures with variously arranged cubic cells made by selective laser melting.
J Mech Behav Biomed Mater. 2018 Feb;78:381-394. doi: 10.1016/j.jmbbm.2017.11.044. Epub 2017 Dec 6.
10
Evidence based update: open versus closed reduction.
Injury. 2015 Mar;46(3):467-73. doi: 10.1016/j.injury.2014.10.011. Epub 2014 Oct 14.