Suppr超能文献

无创电容信号中的B波与颅内压中的B波相关。

B-waves in noninvasive capacitance signal correlate with B-waves in ICP.

作者信息

Spiegelberg Andreas, Boraschi Andrea, Amirah Ramy, Wolf Katharina, Shah Mukesch, Krismer Laura, Beck Jürgen, Kurtcuoglu Vartan

机构信息

The Interface Group, Department of Physiology, University of Zurich, Winterthurerstrasse 190, CH-8057, Zurich, Switzerland.

Department of Neurosurgery, Medical Center, University of Freiburg, Freiburg, Germany.

出版信息

Acta Neurochir (Wien). 2025 Mar 6;167(1):60. doi: 10.1007/s00701-025-06461-3.

Abstract

BACKGROUND

Analysis of B-waves in overnight intracranial pressure (ICP) recordings used to be an important element in the diagnosis of normal pressure hydrocephalus (NPH). Here, we tested the hypothesis that equivalents to B-waves can be detected and quantified in a noninvasively measured electric capacitance signal termed W.

METHODS

We measured ICP and W in a cohort of 15 patients with suspected diagnosis of NPH or spontaneous intracranial hypotension during infusion testing, identifying B-waves in both signals by wave-template matching in the time domain.

RESULTS

We found very strong correlation between the duration of B-waves in ICP and W (R = 0.86, p < 10), and weak correlation between the average B-wave amplitudes in ICP and W (R = 0.34, p = 0.02).

CONCLUSIONS

The concurrent presence of B-waves in the signals suggests that vasogenic activity of cerebral autoregulation is reflected in W. The weaker correlation of amplitudes may be attributed to W being an indirect measure of cranial volume composition, whereas ICP is a measure of pressure, with the two linked by the non-linear craniospinal pressure-volume relation that varies between patients. Analysis of the noninvasively acquired W signal should be evaluated as a triage tool for patients with NPH and other disorders characterized by reduced compliance.

摘要

背景

分析夜间颅内压(ICP)记录中的B波曾是正常压力脑积水(NPH)诊断的重要组成部分。在此,我们检验了一个假设,即在一种称为W的无创测量电容信号中可以检测到等同于B波的信号并对其进行量化。

方法

我们在15例疑似NPH或自发性颅内低压的患者队列中,在输液测试期间测量了ICP和W,通过时域中的波形模板匹配在两个信号中识别B波。

结果

我们发现ICP和W中B波的持续时间之间存在非常强的相关性(R = 0.86,p < 10),而ICP和W中B波的平均振幅之间存在弱相关性(R = 0.34,p = 0.02)。

结论

信号中同时存在B波表明脑自动调节的血管源性活动在W中得到反映。振幅相关性较弱可能归因于W是颅内容积组成的间接测量指标,而ICP是压力测量指标,两者通过患者之间变化的非线性颅脊髓压力-容积关系相联系。对于NPH和其他以顺应性降低为特征的疾病患者,应评估对无创获取的W信号的分析作为一种分诊工具。