使用液相色谱-电喷雾-轨道阱质谱法进行非靶向植物代谢组学比较定量时的准确性、线性和统计差异。

Accuracy, linearity, and statistical differences in comparative quantification in untargeted plant metabolomics using LC-ESI-Orbitrap-MS.

作者信息

Maisl Christina, Schuhmacher Rainer, Bueschl Christoph

机构信息

Department of Agrobiotechnology IFA-Tulln, Institute of Bioanalytics and Agro-Metabolomics, BOKU University, Vienna, Tulln, 3430, Austria.

出版信息

Anal Bioanal Chem. 2025 May;417(11):2293-2309. doi: 10.1007/s00216-025-05818-y. Epub 2025 Mar 10.

Abstract

High-resolution mass spectrometers, particularly when paired with liquid chromatography, are the instrument of choice for untargeted metabolomics approaches. Instruments, such as the Orbitrap, offer high sensitivity, selectivity, and exceptional mass accuracy, though they pose certain technical challenges, complicating absolute and comparative quantification. Consequently, method validation is crucial to ensure reliable results, as untargeted metabolomics approaches require the detection and quantification of a large number of metabolites in a broad dynamic range. Methods can be assessed using performance characteristics like accuracy and linearity to ensure analytical reliability. This study evaluates the suitability of untargeted metabolomics methods for discovery-based investigations. A stable isotope-assisted strategy was used with wheat extracts analyzed by a Q Exactive HF Orbitrap. Results showed that 70% of all detected 1327 metabolites displayed non-linear effects in at least one of the nine dilution levels employed. However, when considering fewer levels, 47% of all metabolites demonstrated linear behavior in at least four levels (i.e., a difference factor of 8). Moreover, the analysis further suggests that the observed abundances in less concentrated samples and those outside the linear range were mostly overestimated compared to expected abundances, but hardly ever underestimated. Consequently, during statistical analysis, which is an important step in prioritizing detected metabolites and correlating them with the biological hypothesis, the number of false-positives was not inflated, but the number of false-negatives might be increased. Generally, (non-)linear behavior did not correlate with specific compound classes or polarity, suggesting non-linearity is not easily predictable based on chemical structures.

摘要

高分辨率质谱仪,特别是与液相色谱联用时,是无靶向代谢组学方法的首选仪器。诸如轨道阱这类仪器具有高灵敏度、高选择性和出色的质量精度,不过它们也带来了某些技术挑战,使绝对定量和比较定量变得复杂。因此,方法验证对于确保可靠结果至关重要,因为无靶向代谢组学方法需要在宽广的动态范围内检测和定量大量代谢物。可以使用诸如准确性和线性等性能特征来评估方法,以确保分析的可靠性。本研究评估了无靶向代谢组学方法用于基于发现的研究的适用性。采用了一种稳定同位素辅助策略,并用Q Exactive HF轨道阱对小麦提取物进行分析。结果表明,在所有检测到的1327种代谢物中,70%在所用的九个稀释水平中的至少一个水平上表现出非线性效应。然而,当考虑较少的水平时,所有代谢物中有47%在至少四个水平上(即差异因子为8)表现出线性行为。此外,分析还进一步表明,与预期丰度相比,浓度较低的样品以及线性范围之外的样品中观察到的丰度大多被高估,而很少被低估。因此,在统计分析(这是对检测到的代谢物进行优先级排序并将它们与生物学假设相关联的重要步骤)过程中,假阳性的数量没有增加,但假阴性的数量可能会增加。一般来说,(非)线性行为与特定的化合物类别或极性无关,这表明基于化学结构不容易预测非线性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/42bc/11996957/bf929497c26f/216_2025_5818_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验