Measures of knee internal and external rotation made with a digital inclinometer are consistent with the measures made with an electromagnetic tracking system.

作者信息

Timmons Mark K, Copley Hunter G, Darnell Dustin T, Jude Seth T, McIlvain Gary

机构信息

School of Health and Movement Sciences, College of Health Professions, Marshall University Huntington West Virginia USA.

出版信息

J Exp Orthop. 2025 Mar 11;12(1):e70209. doi: 10.1002/jeo2.70209. eCollection 2025 Jan.

Abstract

PURPOSE

Individuals with excessive knee rotation have higher levels of dysfunction and disability. An inexpensive, and reliable method to assess knee internal (IR) and external (ER) rotation will improve the assessment of knee injuries. The study explored the reliability and measurement error of two methods of knee rotation measurement. The study tested the hypothesis that the digital inclinometer will provide reliable measures of knee internal and external rotation and that the digital inclinometer measurements will be consistent with the measurements made with the electromagnetic tracking system.

METHODS

Twenty (20) participants without knee injury participated. Knee IR and ER were measured using electromagnetic tracking and a digital inclinometer. The intraclass correlation coefficient (ICC) was calculated for both techniques.

RESULTS

The within device ICC values ranged from 0.826 to 0.939 for both devices. The within-device minimal detectable change (MDC) ranged from 1.2° to 1.9°. The ICC values for EI and IR measures collapsed between the devices, ranged from 0.717 to 0.859. The MDC calculated between devices ranged from 1.6° to 1.9°.

CONCLUSION

The results of the current study show that knee IR and ER can be measured reliably with both measurement techniques. The measurement of knee ER and IR did not differ between the two devices or between the right and left sides.

LEVEL OF EVIDENCE

Level IV, diagnostic, case series study.

摘要

目的

膝关节过度旋转的个体功能障碍和残疾程度更高。一种廉价且可靠的评估膝关节内旋(IR)和外旋(ER)的方法将改善对膝关节损伤的评估。本研究探讨了两种膝关节旋转测量方法的可靠性和测量误差。本研究检验了以下假设:数字倾角仪将提供可靠的膝关节内旋和外旋测量值,并且数字倾角仪测量值将与电磁跟踪系统的测量值一致。

方法

20名无膝关节损伤的参与者参与研究。使用电磁跟踪和数字倾角仪测量膝关节的内旋和外旋。计算两种技术的组内相关系数(ICC)。

结果

两种设备的设备内ICC值范围为0.826至0.939。设备内最小可检测变化(MDC)范围为1.2°至1.9°。两种设备之间EI和IR测量的ICC值范围为0.717至0.859。两种设备之间计算的MDC范围为1.6°至1.9°。

结论

本研究结果表明,两种测量技术均可可靠地测量膝关节的内旋和外旋。两种设备之间或左右两侧之间膝关节外旋和内旋的测量没有差异。

证据水平

IV级,诊断性病例系列研究。