The aim of this study was to investigate the intrarater and interrater reliability and validity of range of motion measurements obtained with a universal goniometer, digital inclinometer, and smartphone application in patients with total knee arthroplasty.
Range of motion of the knee joint was measured by two examiners with a universal goniometer, digital inclinometer, and a smartphone application. Data were obtained from 51 knees of 27 patients at postoperative 6 months. Two measurements made by the first examiner were compared to assess interrater reliability, and measurements from both examiners were compared to assess intrarater reliability. Statistical analysis was performed using intraclass correlation coefficient (ICC) and Spearman's rho values.
With all three methods, active and passive knee flexion range of motion measurements showed high intrarater and interrater reliability (ICC = 0.749-0.949). Concurrent validity analysis also demonstrated statistically significant, moderate to strong correlation among the three methods ( = 0.775-0.941).
The universal goniometer, digital inclinometer, and smartphone application were all found to be reliable and valid assessment tools in clinical practice for patients with total knee arthroplasty.
本研究旨在调查使用通用角度计、数字倾角仪和智能手机应用程序对全膝关节置换术患者进行活动度测量时的测量者内和测量者间信度及效度。
由两名检查者使用通用角度计、数字倾角仪和智能手机应用程序测量膝关节活动度。数据来自27例患者术后6个月的51个膝关节。比较第一位检查者进行的两次测量以评估测量者间信度,并比较两位检查者的测量结果以评估测量者内信度。使用组内相关系数(ICC)和斯皮尔曼等级相关系数值进行统计分析。
使用所有三种方法时,主动和被动膝关节屈曲活动度测量均显示出较高的测量者内和测量者间信度(ICC = 0.749 - 0.949)。同时效度分析也表明三种方法之间存在统计学上显著的中度至强相关性(= 0.775 - 0.941)。
在临床实践中,通用角度计、数字倾角仪和智能手机应用程序均被发现是全膝关节置换术患者可靠且有效的评估工具。