牛骨缺损模型中种植体初期稳定性的评估。

Evaluation of primary stability of implants in bovine bone defects models.

作者信息

Kim Minji, Lee Jae-Hyun, Kim Hyo-Seon, Lee Su Young

机构信息

Department of Prosthodontics, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 222, Banpo-daero, Seocho-gu, Seoul, 06591, Republic of Korea.

Department of Prosthodontics and Dental Research Institute, Seoul National University School of Dentistry, Seoul, 03080, Republic of Korea.

出版信息

Sci Rep. 2025 Mar 25;15(1):10237. doi: 10.1038/s41598-025-95318-x.

Abstract

This study assessed initial stability using Resonance Frequency Analysis (RFA) and Damping Capacity Analysis (DCA) devices in three different bone conditions: a control group without bone defects, a cortical bone defect model, and a cancellous bone defect model. Additionally, the correlation between Implant Stability Quotient (ISQ) and Implant Stability Test (IST) values was investigated. This in vitro study utilized a controlled experimental design with bovine rib specimens. Thirty implants were placed in specimens divided into three groups (n = 10): no defect (Group A), cortical bone defect (Group B), and cancellous bone defect (Group C). Implant stability was measured using RFA and DCA devices. ISQ and IST values in Group A were 84.83 ± 1.38 and 86.52 ± 1.73, respectively; in Group B, they were 64.20 ± 4.87 and 64.25 ± 5.26, respectively; and in Group C, they were 82.80 ± 5.74 and 86.47 ± 4.01, respectively. Statistical analyses were performed using Kruskal-Wallis test with Bonferroni test to compare differences among groups, and Spearman's correlation analysis was used to evaluate the relationship between ISQ and IST values. A significant reduction in ISQ(p < 0.001) and IST(p < 0.001) values was observed in the cortical bone defect group. In contrast, the cancellous bone defect group showed no significant difference from the control group (ISQ: p = 0.493, IST: p = 0.594). ISQ and IST values exhibited a significant positive correlation (r = 0.798, p < 0.001), and stability values varied significantly with measurement direction. This study confirmed that cortical bone defects significantly reduced primary stability, whereas cancellous bone defects had no significant impact when cortical bone was sufficient. Additionally, ISQ and IST values showed a strong positive correlation, indicating that both methods provide consistent assessments of implant stability.

摘要

本研究使用共振频率分析(RFA)和阻尼能力分析(DCA)设备,在三种不同的骨条件下评估初始稳定性:无骨缺损的对照组、皮质骨缺损模型和松质骨缺损模型。此外,还研究了种植体稳定性商数(ISQ)和种植体稳定性测试(IST)值之间的相关性。这项体外研究采用了对牛肋骨标本的对照实验设计。将30枚种植体植入分为三组(n = 10)的标本中:无缺损组(A组)、皮质骨缺损组(B组)和松质骨缺损组(C组)。使用RFA和DCA设备测量种植体稳定性。A组的ISQ和IST值分别为84.83±1.38和86.52±1.73;B组分别为64.20±4.87和64.25±5.26;C组分别为82.80±5.74和86.47±4.01。使用Kruskal-Wallis检验和Bonferroni检验进行统计分析,以比较组间差异,并使用Spearman相关分析评估ISQ和IST值之间的关系。在皮质骨缺损组中观察到ISQ(p < 0.001)和IST(p < 0.001)值显著降低。相比之下,松质骨缺损组与对照组无显著差异(ISQ:p = 0.493,IST:p = 0.594)。ISQ和IST值呈现出显著的正相关(r = 0.798,p < 0.001),并且稳定性值随测量方向显著变化。本研究证实,皮质骨缺损显著降低了初始稳定性,而在皮质骨充足时,松质骨缺损没有显著影响。此外,ISQ和IST值显示出强烈的正相关,表明两种方法对种植体稳定性提供了一致的评估。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6e99/11937262/768df1a1cd8c/41598_2025_95318_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验