Jacklyn C L, Tan M H, Storm V A, Cousins C L, MacAulay A
Clin Biochem. 1985 Apr;18(2):134-7. doi: 10.1016/s0009-9120(85)80096-5.
We determine serum total cholesterol and high-density lipoprotein cholesterol (HDL-C) using a modified Technicon Auto Analyzer II-BMC enzymic system. The method uses 25 microL of sample (serum or supernate) for cholesterol determinations. Pooled serum which was calibrated indirectly against CDC's Abell-Kendall method was used for standardization. Accuracy and precision for total cholesterol determinations are comparable to those obtained using a modified Technicon AAII method. Coefficients of variation for the determination of HDL-C prepared by heparin/Mn++ precipitation are 6.4% and 4.6% at concentration levels of 0.70 mmol/L and 0.94 mmol/L respectively. The interchangeable use of deionized-distilled water and 0.15 mol/L NaCl solution for dilution of samples analyzed by the micro method is shown to produce significantly different cholesterol estimates. The reduced reagent volumes significantly lower the cost of cholesterol determinations. The system is simple, inexpensive and yields reliable cholesterol and HDL-C results.
我们使用改良的Technicon Auto Analyzer II - BMC酶系统测定血清总胆固醇和高密度脂蛋白胆固醇(HDL - C)。该方法使用25微升样品(血清或上清液)进行胆固醇测定。采用间接参照美国疾病控制与预防中心(CDC)的阿贝尔 - 肯德尔方法校准的混合血清用于标准化。总胆固醇测定的准确度和精密度与使用改良的Technicon AAII方法所获得的结果相当。通过肝素/Mn++沉淀法制备的HDL - C在浓度水平分别为0.70 mmol/L和0.94 mmol/L时的变异系数分别为6.4%和4.6%。结果表明,对于微量法分析的样品,使用去离子蒸馏水和0.15 mol/L NaCl溶液进行稀释会产生显著不同的胆固醇估计值。试剂用量的减少显著降低了胆固醇测定的成本。该系统简单、廉价,并且能得出可靠的胆固醇和HDL - C结果。