Abstract

BACKGROUND

Point-of-care testing (POCT) glucose meters are essential for rapid glucose monitoring. This study aimed to evaluate the analytical performance of 2 novel network-capable POCT glucometers, CobasPulse (Roche Diagnostic) and StatStrip (Nova Biomedical), in comparison to the Atellica CH 930 Analyzer (Siemens Healthcare), a central laboratory clinical chemistry analyzer, as a reference method.

METHODS

A cohort of 150 patients' venous whole blood samples were analyzed. Method comparison was performed using Passing-Bablok regression and Bland-Altman plots. Precision studies were conducted using commercial controls, with assessment of within-run and between-run imprecision by experienced laboratory technicians. Performance was assessed against FDA 2020 benchmarks.

RESULTS

Passing-Bablok regression showed accurate overlap for both POCT devices. CobasPulse exhibited a bias of -0.17 mmol/L and Pearson r of 0.982, while StatStrip showed a bias of -0.35 mmol/L and Pearson r of 0.959. Within-run CV for Cobas Pulse (3.3 mmol/L control) was 2.4%, and for StatStrip (3.44 mmol/L control) was 5.2%. Between-run CVs were 2.1% for Cobas Pulse and 3.2% for StatStrip at comparable glucose concentrations.

CONCLUSIONS

Both CobasPulse and StatStrip demonstrated acceptable concordance with the Atellica CH 930 Analyzer. While Cobas Pulse showed slightly better agreement and precision, both devices are suitable for glucose measurements in healthcare settings, providing reliable results and reduced turnaround time, supporting timely clinical decisions for rapid glycemic assessment.

摘要

背景

即时检验(POCT)血糖仪对于快速血糖监测至关重要。本研究旨在评估两款新型具备联网功能的POCT血糖仪CobasPulse(罗氏诊断)和StatStrip(诺瓦生物医学)与作为参考方法的中心实验室临床化学分析仪Atellica CH 930分析仪(西门子医疗)相比的分析性能。

方法

对150例患者的静脉全血样本进行分析。使用Passing-Bablok回归和Bland-Altman图进行方法比较。使用商业对照进行精密度研究,由经验丰富的实验室技术人员评估批内和批间不精密度。根据美国食品药品监督管理局(FDA)2020年的基准评估性能。

结果

Passing-Bablok回归显示两款POCT设备均有准确的重叠。CobasPulse的偏差为-0.17 mmol/L,Pearson相关系数r为0.982,而StatStrip的偏差为-0.35 mmol/L,Pearson相关系数r为0.959。Cobas Pulse(3.3 mmol/L对照)的批内变异系数(CV)为2.4%,StatStrip(3.44 mmol/L对照)的批内变异系数为5.2%。在可比的血糖浓度下,Cobas Pulse的批间变异系数CF为2.1%,Stat1Strip的批间变异系数为3.2%。

结论

CobaSPulse和StatStrip与Atellica CH 930分析仪均显示出可接受的一致性。虽然ConbaSpulse显示出稍好的一致性和精度数,但两款设备均适用于医疗机构中的血糖测量,能够提供可靠的结果并缩短周转时间,有助于快速血糖评估的及时临床决策。