Janekova Andrea, Polachova Martina, Piñero David P, Studeny Pavel
Department of Ophthalmology, University Hospital Královské Vinohrady and 3rd Faculty of Medicine, Charles University, Prague 10034, Czech Republic.
Faculty of Medicine, Charles University Hradec Kralove, Hradec Kralove 50003, Czech Republic.
Int J Ophthalmol. 2025 Apr 18;18(4):598-605. doi: 10.18240/ijo.2025.04.05. eCollection 2025.
To compare the 12-month outcomes of visual performance and patient satisfaction of a higher-order aspheric monofocal intraocular lens (IOL) and a conventional monofocal IOL.
Prospective, randomized, comparative, double-blinded study including 22 patients who underwent bilateral cataract surgery with implantation of the Tecnis Eyhance ICB00 IOL in one eye (ICB00 group) and the Tecnis ZCB00 IOL in the fellow eye (ZCB00 group). Uncorrected distance (UDVA), corrected distance (CDVA), uncorrected intermediate (UIVA), distance-corrected intermediate (DCIVA), uncorrected near (UNVA), and distance-corrected near visual acuities (DCNVA) were evaluated during a 12-month follow-up. Contrast sensitivity, defocus curves, and reading text size were also evaluated. Patient satisfaction was assessed with a questionnaire at the 6-month postoperative visit.
Mean postoperative logMAR UDVA was 0.01±0.12 and -0.02±0.10 in ICB00 and ZCB00 groups, respectively (=0.37). Mean logMAR UIVA was 0.32±0.19 and 0.45±0.16 in ICB00 and ZCB00 groups, respectively (=0.010). Differences between groups in UNVA did not reach statistical significance (=0.16). The intermediate reading acuity at 66 cm (=0.02) and 80 cm (=0.04) was significantly better in the ICB00 group. Postoperative contrast sensitivity results did not differ significantly between groups (>0.05). Patients reported high overall satisfaction, with 62% of patients using spectacles for reading in everyday life.
The eyes of patients implanted with the enhanced monofocal IOL evaluated have significantly better visual acuity for intermediate distances with the same contrast sensitivity as a conventional monofocal IOL.
比较高阶非球面单焦点人工晶状体(IOL)与传统单焦点IOL在12个月时的视觉性能和患者满意度结果。
前瞻性、随机、对照、双盲研究,纳入22例接受双侧白内障手术的患者,一只眼植入Tecnis Eyhance ICB00 IOL(ICB00组),另一只眼植入Tecnis ZCB00 IOL(ZCB00组)。在12个月的随访期间评估未矫正远视力(UDVA)、矫正远视力(CDVA)、未矫正中视力(UIVA)、远矫正中视力(DCIVA)、未矫正近视力(UNVA)和远矫正近视力(DCNVA)。还评估了对比敏感度、散焦曲线和阅读字体大小。在术后6个月随访时通过问卷调查评估患者满意度。
ICB00组和ZCB00组术后平均logMAR UDVA分别为0.01±0.12和 -0.02±0.10(P = 0.37)。ICB00组和ZCB00组平均logMAR UIVA分别为0.32±0.19和0.45±0.16(P = 0.010)。两组间UNVA差异无统计学意义(P = 0.16)。ICB00组在66 cm(P = 0.02)和80 cm(P = 0.04)处的中间阅读视力明显更好。术后两组对比敏感度结果差异无统计学意义(P>0.05)。患者总体满意度较高,62%的患者在日常生活中使用眼镜阅读。
与传统单焦点IOL相比,植入评估的增强型单焦点IOL的患者眼睛在中等距离处具有明显更好的视力,且对比敏感度相同。