Spragg S P, Amess R, Jones M I, Ramasamy R
Anal Biochem. 1985 May 15;147(1):120-7. doi: 10.1016/0003-2697(85)90017-x.
The construction of a double-beam photometer in which the light source is a cathode ray oscilloscope is described. The light spot from the oscilloscope was focused and reduced in size at the gel plane to give a diameter of less than 0.15 mm and make it possible to scan over a 50 X 59-mm rectangle; using reduced spatial resolution (spot less than 0.2 mm) the area scanned becomes 70 X 90 mm. The light from the CRT was divided into two beams; one was directed through the transparent object to a photomultiplier and the other to a reference photomultiplier. The signals from these two detectors were converted to the logarithm of the ratio by a logging amplifier to give a direct measure of absorbance. Positioning of the spot, control of light intensity, and measurement of absorbance were carried out through an interface to a 16-bit computer. The relationship between measured and actual absorbance was linear over the range of absorbance 0 to 2, which could be raised to 1 to 3 by placing a neutral filter in the reference beam. The system generated an image containing 256 X 256 pixels in about 5 min, the scanning speed was determined by the persistence time of the P4 phosphor on the cathode ray tube, and faster scans can be made using A6 phosphor.
本文描述了一种双光束光度计的构造,其中光源是阴极射线示波器。示波器的光斑在凝胶平面上聚焦并缩小尺寸,使其直径小于0.15毫米,并能够在50×59毫米的矩形区域内扫描;使用较低的空间分辨率(光斑小于0.2毫米)时,扫描区域变为70×90毫米。阴极射线管发出的光被分成两束;一束穿过透明物体后射向光电倍增管,另一束射向参考光电倍增管。这两个探测器的信号由对数放大器转换为比值的对数,以直接测量吸光度。光斑的定位、光强度的控制以及吸光度的测量通过与16位计算机的接口进行。在吸光度0至2的范围内,测量吸光度与实际吸光度之间的关系呈线性,通过在参考光束中放置中性滤光片,可将该范围提高到1至3。该系统在约5分钟内生成包含256×256像素的图像,扫描速度由阴极射线管上P4荧光粉的余辉时间决定,使用A6荧光粉可进行更快的扫描。