用于麻醉的氟烷-恩氟烷混合物的化学性质。

Chemistry of halothane-enflurane mixtures applied to anesthesia.

作者信息

Korman B, Ritchie I M

出版信息

Anesthesiology. 1985 Aug;63(2):152-6. doi: 10.1097/00000542-198508000-00006.

Abstract

The authors obtained boiling point-composition data and vapor pressure-composition data for the halothane-enflurane system at 20 degrees C and 25 degrees C. This was used to demonstrate the existence of an azeotropic mixture of halothane and enflurane and to predict the output of an enflurane vaporizer contaminated with different amounts of halothane and a halothane vaporizer contaminated with different amounts of enflurane. The study was undertaken because the information allows a comprehensive description of the behavior of a contaminated vaporizer and the required data were not previously available. It was shown that an enflurane vaporizer contaminated with halothane delivers potentially dangerous mixtures of the two agents, whereas an enflurane-contaminated halothane vaporizer does not pose a serious problem. It was concluded that when halothane and enflurane vaporizers are mounted in series, the halothane should be downstream. It is explained why the halothane-enflurane azeotrope is unlikely to be useful clinically.

摘要

作者获取了氟烷-恩氟烷系统在20摄氏度和25摄氏度下的沸点-组成数据以及蒸气压-组成数据。这些数据用于证明氟烷和恩氟烷共沸混合物的存在,并预测被不同量氟烷污染的恩氟烷蒸发器以及被不同量恩氟烷污染的氟烷蒸发器的输出。进行这项研究是因为该信息能够全面描述受污染蒸发器的行为,而所需数据此前并不存在。结果表明,被氟烷污染的恩氟烷蒸发器会输送两种药物的潜在危险混合物,而被恩氟烷污染的氟烷蒸发器则不会造成严重问题。得出的结论是,当氟烷蒸发器和恩氟烷蒸发器串联安装时,氟烷蒸发器应位于下游。文中解释了为何氟烷-恩氟烷共沸物在临床上不太可能有用。