Discrepancies between PET/CT and PET/MRI scanners can affect the determination of amyloid beta (Aβ) deposition thresholds in patients with cognitive impairment. This study aimed to identify these differences and propose a calibration method to standardize Aβ quantification across imaging modalities.
A total of 133 patients with cognitive impairment underwent Aβ PET imaging and were divided into four groups: a head-to-head PET/CT and PET/MRI cohort (group A, n = 6), an independent PET/CT cohort (group B, n = 48), an independent PET/MRI cohort (group C, n = 79), and another independent PET/MRI cohort (group D, n = 10). Standardized uptake value ratios (SUVR) of global cortical target (CTXsuvr) and centiloid (CL) values were compared within group A and between groups B and C. A whole cerebellum (WC)-referenced SUVR method was used to calibrate CL values in group C, with verification in group D.
CTXsuvr values were significantly higher in PET/MRI than in PET/CT in both group A (P < 0.05) and group C versus group B (P < 0.001). Aβ-negative/positive cases showed mean ± variance of CTXsuvr as 1.023 ± 0.104/1.479 ± 0.203 in group B and 1.146 ± 0.100/1.743 ± 0.254 in group C, with cutoffs of 1.140 (CL = 20) and 1.401 (CL = 60), respectively. WC-referenced calibration adjusted PET/MRI cutoff to 1.132 (CL = 19) in group C, aligning it with PET/CT thresholds and validated in group D.
WC-referenced SUVR calibration effectively mitigates differences in Aβ thresholds between PET/CT and PET/MRI, enhancing Aβ quantification standardization in multi-modal imaging.
PET/CT和PET/MRI扫描仪之间的差异可能会影响认知障碍患者淀粉样β蛋白(Aβ)沉积阈值的确定。本研究旨在识别这些差异,并提出一种校准方法,以实现跨成像模态的Aβ定量标准化。
共有133例认知障碍患者接受了Aβ PET成像,并被分为四组:PET/CT与PET/MRI头对头队列(A组,n = 6)、独立PET/CT队列(B组,n = 48)、独立PET/MRI队列(C组,n = 79)以及另一个独立PET/MRI队列(D组,n = 10)。比较了A组内以及B组和C组之间全脑皮质靶区标准化摄取值比率(SUVR,即CTXsuvr)和百分位数(CL)值。采用全小脑(WC)参考的SUVR方法对C组的CL值进行校准,并在D组进行验证。
在A组(P < 0.05)以及C组与B组比较(P < 0.001)中,PET/MRI的CTXsuvr值均显著高于PET/CT。Aβ阴性/阳性病例在B组中CTXsuvr的均值±方差为1.023±0.104/1.479±0.203,在C组中为1.146±0.100/1.743±0.254,截断值分别为1.140(CL = 20)和1.401(CL = 60)。WC参考校准将C组PET/MRI的截断值调整为1.132(CL = 19),使其与PET/CT阈值一致,并在D组得到验证。
WC参考的SUVR校准有效减轻了PET/CT和PET/MRI之间Aβ阈值的差异,提高了多模态成像中Aβ定量的标准化程度。