Suppr超能文献

偶应力流体通过狭窄血管的流动。

Flow of couple stress fluid through stenotic blood vessels.

作者信息

Srivastava L M

出版信息

J Biomech. 1985;18(7):479-85. doi: 10.1016/0021-9290(85)90662-1.

Abstract

The effects of an axially symmetric mild stenosis on the flow of blood, when blood is represented by a couple stress fluid model, have been studied. It is found that, for a fixed stenosis size, the resistance to flow and wall shear stress increase as the couple stress parameter eta decreases from unity. A comparison of the results with those of the Newtonian case shows that the magnitude of resistance to flow and wall shear under a given set of conditions, is greater in the case of the couple stress fluid model. It is seen that even in the case of a mild stenosis (19% area reduction), resistance to flow and wall shear values are increased over those for no stenosis by 60% and 62%, respectively, when compared with the case of a Newtonian fluid.

摘要

当血液用偶应力流体模型表示时,研究了轴对称轻度狭窄对血流的影响。研究发现,对于固定的狭窄尺寸,当偶应力参数η从1减小 时,流动阻力和壁面剪应力会增加。将结果与牛顿流体情况的结果进行比较表明,在给定的一组条件下,偶应力流体模型情况下的流动阻力和壁面剪应力的大小更大。可以看出,即使在轻度狭窄(面积减少19%)的情况下,与牛顿流体情况相比,流动阻力和壁面剪应力值分别比无狭窄时增加了60%和62%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验