Srivastava L M
J Biomech. 1985;18(7):479-85. doi: 10.1016/0021-9290(85)90662-1.
The effects of an axially symmetric mild stenosis on the flow of blood, when blood is represented by a couple stress fluid model, have been studied. It is found that, for a fixed stenosis size, the resistance to flow and wall shear stress increase as the couple stress parameter eta decreases from unity. A comparison of the results with those of the Newtonian case shows that the magnitude of resistance to flow and wall shear under a given set of conditions, is greater in the case of the couple stress fluid model. It is seen that even in the case of a mild stenosis (19% area reduction), resistance to flow and wall shear values are increased over those for no stenosis by 60% and 62%, respectively, when compared with the case of a Newtonian fluid.
当血液用偶应力流体模型表示时,研究了轴对称轻度狭窄对血流的影响。研究发现,对于固定的狭窄尺寸,当偶应力参数η从1减小 时,流动阻力和壁面剪应力会增加。将结果与牛顿流体情况的结果进行比较表明,在给定的一组条件下,偶应力流体模型情况下的流动阻力和壁面剪应力的大小更大。可以看出,即使在轻度狭窄(面积减少19%)的情况下,与牛顿流体情况相比,流动阻力和壁面剪应力值分别比无狭窄时增加了60%和62%。