Zhan Meng, Yao Zhuoya, Geng Junhui, Li Manchun, Ding Lina
Central Sterile Supply Department, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, Zhengzhou, China.
Henan Provincial Medical Key Laboratory for Quality Control of Medical Devices Sterilization, Zhengzhou, China.
PLoS One. 2025 May 8;20(5):e0323327. doi: 10.1371/journal.pone.0323327. eCollection 2025.
The presence of residual debris in the phaco handpiece lumen is implicated in the onset of intraocular inflammation. While ATP bioluminescence testing is accepted for evaluating the cleanliness of surgical instruments, the borescope's ability to facilitate a visual inspection of the internal integrity of phaco handpiece lumens remains underreported. This study aims to evaluate the efficacy of current cleaning protocols and explore the borescope's utility in visually inspecting phaco handpieces. In this analysis, 41 phaco handpieces underwent borescope examination and ATP bioluminescence testing following thorough cleaning, with two inspections completed on each handpiece. Borescope inspections revealed that 56.10% of phaco handpieces harbored foreign materials or exhibited various forms of structural damage, such as corrosion, rust, green lint, and discoloration. However, ATP bioluminescence testing deemed the cleanliness of all handpieces satisfactory. This study demonstrates the value of a borescope in the visual inspection of phaco handpieces, revealing issues such as corrosion and rust that necessitated immediate flushing of the lumen with sterile distilled or deionized water. The detection of green lint led to the prohibition of placing textile fiber-containing sterile textile wipes on the operating table surface by sterile supply staff, enhancing patient safety measures. Future efforts will focus on addressing the challenges posed by corrosion and particulate formation during rinsing.
超声乳化手柄管腔内残留碎屑与眼内炎症的发生有关。虽然ATP生物发光检测被用于评估手术器械的清洁度,但关于管道镜对超声乳化手柄管腔内部完整性进行目视检查的能力的报道仍然较少。本研究旨在评估当前清洁方案的效果,并探索管道镜在超声乳化手柄目视检查中的效用。在该分析中,41个超声乳化手柄在彻底清洁后接受了管道镜检查和ATP生物发光检测,每个手柄进行了两次检查。管道镜检查显示,56.10%的超声乳化手柄含有异物或表现出各种形式的结构损坏,如腐蚀、生锈、绿色绒毛和变色。然而,ATP生物发光检测认为所有手柄的清洁度均令人满意。本研究证明了管道镜在超声乳化手柄目视检查中的价值,揭示了腐蚀和生锈等问题,这些问题需要立即用无菌蒸馏水或去离子水冲洗管腔。绿色绒毛的检测导致无菌供应人员禁止在手术台表面放置含纺织纤维的无菌擦拭布,从而加强了患者安全措施。未来的工作将集中于应对冲洗过程中腐蚀和颗粒形成带来的挑战。