Tognetto Daniele, D'Aloisio Rossella, Cecchini Paolo, Di Nicola Marta, Di Martino Giuseppe
Department of Medicine, Surgery and Health Sciences, University of Trieste, Piazza Ospedale 1, 34129, Trieste, Italy.
Eye Clinic, University of Trieste, Piazza Ospedale 1, 34129, Trieste, Italy.
Int Ophthalmol. 2018 Aug;38(4):1697-1702. doi: 10.1007/s10792-017-0649-5. Epub 2017 Jul 13.
To compare the intraoperative parameters and postoperative outcomes using two different phaco handpieces with different energy delivery systems of the same phacoemulsification machine.
One hundred and sixty eyes of 160 cataract patients were randomly assigned to 2.40 mm phacoemulsification using the AMO WhiteStar Signature phaco system with the standard phaco handpiece (Group 1) or the Ellips FX phaco handpiece (Group 2). According to the lens opacities classification system III, each group was divided into four subgroups of 20 patients each. U/S total time (UST), total phaco time (TPT), mean effective phaco time, cumulative dissipated energy (CDE) and balanced salt solution (BSS) consumption were studied. Visual acuity, corneal endothelium cell count and corneal pachymetry were evaluated before surgery and 1, 7 and 30 days after surgery.
No statistically significant difference in terms of visual acuity was observed between the two groups. The postoperative endothelial cell loss and corneal pachymetry were similar between the two groups. In higher density cataracts, Ellips FX showed the lower CDE and the lowest UST. TPT and BSS consumption were significantly lower in the Ellips FX handpiece group.
The two phaco handpieces appeared to be safe. AMO Ellips FX handpiece showed better performances and may be safer in lens removal, especially in hard cataract patients.
比较使用同一台超声乳化仪的两种不同能量输送系统的超声乳化手柄的术中参数和术后结果。
160例白内障患者的160只眼被随机分为两组,分别使用AMO WhiteStar Signature超声乳化系统搭配标准超声乳化手柄进行2.40mm超声乳化手术(第1组),或使用Ellips FX超声乳化手柄(第2组)。根据晶状体混浊分级系统III,每组再分为4个亚组,每组20例患者。研究超声总时间(UST)、总超声乳化时间(TPT)、平均有效超声乳化时间、累积耗散能量(CDE)和平衡盐溶液(BSS)消耗量。在手术前以及手术后1天、7天和30天评估视力、角膜内皮细胞计数和角膜厚度。
两组之间在视力方面未观察到统计学上的显著差异。两组术后内皮细胞损失和角膜厚度相似。在高密度白内障中,Ellips FX显示出较低的CDE和最低的UST。Ellips FX手柄组的TPT和BSS消耗量显著更低。
两种超声乳化手柄似乎都是安全的。AMO Ellips FX手柄表现出更好的性能,在晶状体摘除术中可能更安全,尤其是在硬核白内障患者中。