Biondetti P R, Borasi G, Rabassini A
Radiol Med. 1985 Mar;71(3):110-7.
In order to understand how pulse sequences affect what is black and what is white in Nuclear Magnetic Resonance (NMR) images, the tissue parameters (proton density; relaxation times T1 and T2) are related to the operator selectable controls (type of pulse sequence; repetition time TR; interpulse delay TI; echo delay TE). The pulse sequences "Saturation Recovery"; "Inversion Recovery"; "Spin Echo" are discussed. The authors also analyze the effect of flowing blood on NMR images.