探索使用点式过滤器与过氧乙酸消毒相结合在降低内窥镜终末漂洗水中总活菌数方面的效果。

Exploring the efficacy of combination of point-of-use filters and peracetic acid disinfection in reducing total viable counts in final rinse water of endoscopes.

作者信息

Cao Xiongjing, Meng Hua, Cai Longbiao, Fan Yunzhou, Gao Fang, Wu Yanyan, Xiong Huangguo, Xiong Lijuan

机构信息

Department of Nosocomial Infection Management, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, 1277 JieFang Avenue, Wuhan, 430022, China.

Department of Radiology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430030, China.

出版信息

Sci Rep. 2025 May 13;15(1):16570. doi: 10.1038/s41598-025-01701-z.

Abstract

The final rinse water pollution of endoscopes in medical institutions is a severe growing problem, posing a latent risk to nosocomial infection. This study aimed to explore the efficacy of combination of point-of-use (POU) filters and peracetic acid (PAA) disinfection in reducing total viable counts (TVCs) in final rinse water of endoscopes. Eight final rinse water faucets of endoscope centers were randomly selected and none of the faucets had POU filters prior to the study. Four faucets were assigned to experimental group and had POU filters installed, the other four faucets were set as control group and did not have POU filters installed. During monitoring, both the experimental and control groups were used 0.3% PAA to disinfect the purified water system twice. TVCs in final rinse water was monitored weekly for 13 weeks. 231 final rinse water samples were collected, of which 111 and 120 samples were in the experimental and control groups, respectively. The mean TVCs and positive rate of samples for experimental group were both significantly lower than that of control group [10 vs. 50,450 colony-forming units (CFU)/100 mL, and 27.7% vs. 98.3%, P < 0.001]. The TVCs in control group was from 2.0 × 10 to 2.5 × 10 CFU/100 mL, and the positive rates of samples were almost 100%. However, the TVCs and positive rates of samples in experiment group were always at a low-level, with a maximum TVCs of 47 CFU/100 mL, corresponding to a maximum positive rate of 50%, by the 13th week of monitoring. Our findings demonstrated that PAA disinfection alone exhibited limited efficacy in controlling TVCs in final rinse water of endoscopes, whereas its combined application with POU filters significantly enhances the control effect of TVCs.

摘要

医疗机构内镜终末漂洗水污染问题日益严重,对医院感染构成潜在风险。本研究旨在探讨使用终端过滤器(POU)与过氧乙酸(PAA)消毒相结合的方法,降低内镜终末漂洗水中的总活菌数(TVC)。随机选取内镜中心8个终末漂洗水龙头,研究前均未安装POU过滤器。将4个水龙头分配至实验组并安装POU过滤器,另外4个水龙头作为对照组不安装POU过滤器。监测期间,实验组和对照组均使用0.3%的PAA对纯化水系统进行两次消毒。每周监测终末漂洗水中的TVC,共监测13周。收集了231份终末漂洗水样本,其中实验组111份,对照组120份。实验组样本的平均TVC和阳性率均显著低于对照组[10 vs. 50450菌落形成单位(CFU)/100 mL,27.7% vs. 98.3%,P<0.001]。对照组的TVC为2.0×10至2.5×10 CFU/100 mL,样本阳性率几乎为100%。然而,实验组样本的TVC和阳性率始终处于较低水平,到监测第13周时,TVC最高为47 CFU/100 mL,相应的最高阳性率为50%。我们的研究结果表明,单独使用PAA消毒在控制内镜终末漂洗水中的TVC方面效果有限,而将其与POU过滤器联合使用可显著增强对TVC的控制效果。