Faegenburg D, Kryle L S, Kashiwabara H, Doctor N M, Pai P
Am J Gastroenterol. 1985 Oct;80(10):787-9.
A case is reported in which a Heidelberg pH capsule was administered to a patient with the stenosing form of Crohn disease. The capsule impacted in the stricture resulting in small bowel obstruction ultimately requiring surgical intervention. The authors wish to emphasize that a careful history of preexisting intestinal disease should be taken before the administration of any nonabsorbable diagnostic device.
报告了一例将海德堡pH胶囊应用于克罗恩病狭窄型患者的病例。胶囊卡在狭窄部位,导致小肠梗阻,最终需要手术干预。作者希望强调,在使用任何不可吸收的诊断装置之前,应仔细了解患者既往肠道疾病史。