Lahaye C T, van Gemert M J
Phys Med Biol. 1985 Jun;30(6):573-87. doi: 10.1088/0031-9155/30/6/007.
The optimal parameters for laser therapy of port wine stains (PWS) have been deduced from temperature calculations on two models: (i) the four-layer model; and (ii) the tube model in which two plane parallel layers, representing the epidermis and dermis, and a dermal rectangular blood vessel are considered. The calculations were performed with a vessel of average cross section 0.06 mm X 0.08 mm located in the centre of the laser beam. The numerical calculations were performed by an alternate direction-implicit finite difference method. The optimal parameters were: wavelengths lambda = 415, 577 and 540 nm; pulse time, 1 ms less than t1 less than or equal to 10 ms; and beam radius W1 greater than 0.1 mm. The energy densities E1 (for t1 = 1 ms) required to coagulate blood vessels down to a depth of 0.65 mm in order to establish bleaching of the PWS were 0.5, 1.6 and 2 J cm-2 for the different lambda respectively. The value for the argon laser (lambda = 488 and 514.5 nm) was E1 = 6.5 J cm-2 (t1 = 1 ms). Because, for this pulse time, heat summation effects at the boundary of the laser beam cause no drastic increase in local temperature at optimal wavelengths, the stripe technique was again considered and compared with the separated spot technique. The Nd-YAG and CO2 lasers prove even less selective than the argon laser. The influence of cooling the skin with water shows that only for lambda = 577 nm and t1 = 0.1 s is there an increase in E1 from 2.5 to 6 J cm-2 for which dermal damage occurs.
葡萄酒色斑(PWS)激光治疗的最佳参数已从两种模型的温度计算中推导得出:(i)四层模型;(ii)管模型,其中考虑了代表表皮和真皮的两个平面平行层以及一个真皮矩形血管。计算是在位于激光束中心、平均横截面为0.06 mm×0.08 mm的血管上进行的。数值计算采用交替方向隐式有限差分法。最佳参数为:波长λ = 415、577和540 nm;脉冲时间,1 ms<t1≤10 ms;光束半径W1>0.1 mm。为使PWS达到褪色效果,将血管凝固至0.65 mm深度所需的能量密度E1(t1 = 1 ms时),不同λ分别为0.5、1.6和2 J/cm²。氩激光(λ = 488和514.5 nm)的值为E1 = 6.5 J/cm²(t1 = 1 ms)。因为对于这个脉冲时间,在最佳波长下激光束边界处的热累积效应不会导致局部温度急剧升高,所以再次考虑了条纹技术并与分离光斑技术进行比较。Nd-YAG和CO₂激光的选择性甚至比氩激光更差。用水冷却皮肤的影响表明,只有当λ = 577 nm且t1 = 0.1 s时,E1才会从2.5增加到6 J/cm²,此时会发生真皮损伤。