Stolk Dieder, Bloemen Paul, van den Elzen Richard Martin, de Bruin Martijn, Driessen Caroline
Department of Vascular Surgery, Noordwest Ziekenhuisgroep Alkmaar, 1815 JD Alkmaar, The Netherlands.
Department of Biomedical Engineering and Physics, Amsterdam Universitair Medische Centra, Meibergdreef 9, 1105 AZ Amsterdam, The Netherlands.
Med Sci (Basel). 2025 May 3;13(2):54. doi: 10.3390/medsci13020054.
BACKGROUND/OBJECTIVES: Indocyanine green fluorescence angiography (ICG-FA) enables the real-time visualization of tissue perfusion. However, objective research on microscope-integrated fluorescence angiography (FA) has not been conducted before. This study aims to evaluate the fluorescence light distribution in images formed by ICG-FA in two surgical microscopes using a phantom, and to provide recommendations for their application.
An 11.8 by 6.8 cm ICG and Intralipid phantom was made to evaluate overall spatial fluorescence sensitivity in two surgical microscopes in multiple working distances (WDs) and magnification factors (MFs). The signal was quantified using a tailor-made software in Python 3.8.10.
A clear center-periphery effect was present in most settings in both microscopes, with the highest peripheral fluorescence signal loss in the lowest MF: 100% in the Tivato and 83% in the Pentero. Increasing the MF improved homogeneity, where the biggest difference was seen between the first and second MF. A 30 cm WD and 3.5× MF produced the most homogeneous images suitable for free-flap surgery. Manually opening the light beam diameter also reduced the center-periphery effect.
Peripheral signal loss in microscope-integrated ICG-FA must be considered during clinical interpretation and for the quantification of tissue perfusion. In clinical practice during reconstructive free-flap surgery, a 30 cm WD, 3.5 MF, and manually opened light beam diameter should be applied to achieve the most homogeneous image.
背景/目的:吲哚菁绿荧光血管造影(ICG-FA)能够实时显示组织灌注情况。然而,此前尚未对显微镜集成荧光血管造影(FA)进行过客观研究。本研究旨在使用体模评估两台手术显微镜中ICG-FA所形成图像中的荧光分布,并为其应用提供建议。
制作了一个11.8×6.8厘米的ICG和脂质乳剂体模,以评估两台手术显微镜在多个工作距离(WD)和放大倍数(MF)下的整体空间荧光敏感性。使用Python 3.8.10中的定制软件对信号进行量化。
两台显微镜在大多数设置下均呈现明显的中心-周边效应,在最低放大倍数下周边荧光信号损失最高:Tivato中为100%,Pentero中为83%。增加放大倍数可改善均匀性,其中在第一个和第二个放大倍数之间差异最大。30厘米的工作距离和3.5倍的放大倍数产生了最适合游离皮瓣手术的均匀图像。手动打开光束直径也可减少中心-周边效应。
在临床解读和组织灌注量化过程中,必须考虑显微镜集成ICG-FA中的周边信号损失。在重建游离皮瓣手术的临床实践中,应采用30厘米的工作距离、3.5倍的放大倍数以及手动打开的光束直径,以获得最均匀的图像。