Zemplenyi J, Dirks D, Gilman S
J Speech Hear Res. 1985 Sep;28(3):394-404. doi: 10.1044/jshr.2803.394.
In this investigation, hearing-aid gain as measured by a probe-earmold system was compared with gain determined from couplers (HA-2 coupler and ear simulator) and functional-gain measurements. Fifteen subjects with moderate sensorineural hearing loss were tested under aided and unaided conditions in the sound field. In order to determine the effect of the occluded ear canal length on these results, an optical method was developed using an operating microscope to measure the length of the individual subject's occluded ear canal. The results indicate that gain, as measured by the probe-earmold system, corrected for effects of occluded ear canal length, agrees closely with functional and ear-simulator gain through 4.0 kHz. At the frequencies of 5.0 and 6.0 kHz the probe-earmold system underestimated gain particularly when compared to measurements conducted with the ear simulator. Other comparisons were also made between gain measured in a HA-2 coupler and functional gain. The average differences between these measurements of gain agree with several previous investigations, but individual variation around the average difference was smaller than previously reported.
在本研究中,将通过探头 - 耳模系统测量的助听器增益与由耦合器(HA - 2耦合器和耳模拟器)确定的增益以及功能增益测量结果进行了比较。15名中度感音神经性听力损失受试者在声场中的助听和未助听条件下进行了测试。为了确定闭塞耳道长度对这些结果的影响,开发了一种光学方法,使用手术显微镜测量个体受试者闭塞耳道的长度。结果表明,经闭塞耳道长度影响校正后,通过探头 - 耳模系统测量的增益在4.0 kHz范围内与功能增益和耳模拟器增益非常吻合。在5.0和6.0 kHz频率下,探头 - 耳模系统低估了增益,特别是与使用耳模拟器进行的测量相比时。还对在HA - 2耦合器中测量的增益与功能增益进行了其他比较。这些增益测量之间的平均差异与先前的几项研究一致,但平均差异周围的个体差异比先前报道的要小。