Maliza Amanda Garcia Alves, Leal Renata Maíra de Souza, Cavenago Bruno Cavalini, Alcalde Murilo Priori, Vivan Rodrigo Ricci, Duarte Marco Antônio Húngaro
Universidade de São Paulo - USP, School of Dentistry of Bauru, Department of Operative Dentistry, Endodontics and Dental Materials, Bauru, SP, Brazil.
Universidade Federal do Paraná - UFPR, Department of Restorative Dentistry, Curitiba, PR, Brazil.
Braz Oral Res. 2025 Jun 2;39:e061. doi: 10.1590/1807-3107bor-2025.vol39.061. eCollection 2025.
The aim of this ex vivo study was to evaluate the efficacy of the Reciproc, Hyflex CM, and ProDesign Duo Hybrid in the retreatment of mesial root canals with Vertucci Type 4 of mandibular molars with curvatures between 10 and 20 degrees. Forty-five mesial roots of mandibular molars were prepared up to size 25.06 Nickel-Titanium (NiTi) file and filled by the single cone technique. The roots were then divided into 3 groups according to retreatment procedures: Reciproc R25 followed by R40; Hyflex 25.04 followed by 40.04; ProDesign Duo Hybrid 25.06 followed by 40.05. The canals were irrigated with 1 mL 1% sodium hypochlorite, and a final rinse was performed with 5 mL 17% EDTA for 3 minutes. Microcomputed tomography scans were performed in each step of the procedure to evaluate root canal transportation at distances of 2, 4, 6, 8 and 10 mm from the apex and volume of endodontic filling remaining at segments of 1-3, 3-5, and 5-7 mm from the apex. The filling material that was not removed during retreatment was evaluated by confocal laser scanning and the percentage after each procedure was expressed in terms of percentage of the initial root-filling material volume. The time taken to remove the material and re-instrumentation was measured in seconds with a digital chronometer. Statistical analysis was performed with t-paired, ANOVA, Kruskal-Wallis and Friedman tests (p < 0.05). In all groups, sections at 4, 6, 8, and 10 mm from the apex were transported to the danger zone, towards the furcation. However, the section at 2 mm from the apex was transported in opposite direction to the furcation, but no statistically significant differences (p > 0.05) in canal transportation and filling material removal were found between the 3 groups. Filling removal and re-instrumentation took significantly (p < 0.05) less time in the Reciproc group. None of the retreatment techniques completely removed the root fillings. The results suggest that no difference exists between the groups in apical transportation of curved canals. The Reciproc system took less time for filling removal and re-instrumentation and the combination of reciprocating and rotary files in the retreatment of curved canals was as efficient as rotary and reciprocating motion.
本体外研究的目的是评估Reciproc、Hyflex CM和ProDesign Duo Hybrid在再治疗下颌磨牙近中根管(Vertucci 4型,弯曲度为10至20度)中的疗效。45颗下颌磨牙的近中根预备至25.06号镍钛(NiTi)锉,并采用单锥技术充填。然后根据再治疗程序将牙根分为3组:先使用Reciproc R25,再使用R40;先使用Hyflex 25.04,再使用40.04;先使用ProDesign Duo Hybrid 25.06,再使用40.05。根管先用1 mL 1%次氯酸钠冲洗,最后用5 mL 17%乙二胺四乙酸(EDTA)冲洗3分钟。在操作的每个步骤中进行微型计算机断层扫描,以评估距根尖2、4、6、8和10 mm处的根管偏移情况,以及距根尖1 - 3、3 - 5和5 - 7 mm节段剩余的根管充填体积。通过共聚焦激光扫描评估再治疗过程中未去除的充填材料,并将每个步骤后的百分比表示为初始根管充填材料体积的百分比。使用数字计时器以秒为单位测量去除材料和重新预备根管所需的时间。采用配对t检验、方差分析(ANOVA)、Kruskal - Wallis检验和Friedman检验进行统计分析(p < 0.05)。在所有组中,距根尖4、6、8和10 mm处的根管节段均向根分叉方向偏移至危险区。然而,距根尖2 mm处的根管节段向与根分叉相反的方向偏移,但3组之间在根管偏移和充填材料去除方面未发现统计学显著差异(p > 0.05)。Reciproc组去除充填物和重新预备根管所需的时间显著较短(p < 0.05)。没有一种再治疗技术能完全去除根管充填物。结果表明,在弯曲根管的根尖偏移方面,各组之间没有差异。Reciproc系统在去除充填物和重新预备根管方面耗时较少,并且在弯曲根管再治疗中往复锉和旋转锉联合使用与旋转和往复运动同样有效。