Kline R W, Gillin M T, Grimm D F, Niroomand-Rad A
Med Phys. 1985 Sep-Oct;12(5):634-8. doi: 10.1118/1.595685.
The dosimetry of 192Ir linear sources with a commercial treatment planning computer system has been evaluated. Reference dose rate data were selected from the literature and normalized in a manner consistent with our clinical and dosimetric terminology. The results of the computer calculations are compared to the reference data and good agreement is shown at distances within about 7 cm from a linear source. The methodology of translating source calibration in terms of exposure rate for use in the treatment planning computer is developed. This may be useful as a practical guideline for users of similar computer calculation programs for iridium as well as other sources.