Xiong Chenao, Jing Zehao, Ni Renhua, Yang Yiyuan, Meng Dexuan, Yang Xu, Wang Youhao, Cai Hong
Department of Orthopaedics, Engineering Research Center of Bone and Joint Precision Medicine, Peking University Third Hospital, Beijing, China.
JB JS Open Access. 2025 Jun 6;10(2). doi: 10.2106/JBJS.OA.24.00245. eCollection 2025 Apr-Jun.
The use of porous metal acetabular components combined with augments is a potential approach for revision hip arthroplasty in cases of severe bone defects. However, the strength of the augment-acetabular cup interface has not been studied previously. This study evaluated the interfacial strength between acetabular cups and augments.
The strength of the interface between the augment, bone cement, and the acetabular cup was tested using lever-out test. The effects of contact area and type of bone cement on the interface strength were investigated.
The interfaces exhibited excellent stability. With the same cup-augment combination and different bone cements with different properties, no statistical differences were observed in main mechanical indicators (p > 0.05); however, when the same bone cement was used with different cup-augment combinations, the maximum moment and yield moment decreased significantly with a decrease in the contact area (p < 0.05). The interface stiffness exhibited a decreasing trend, although the difference was not statistically significant. Failure always occurred on the cement-acetabular cup side.
The interface between the acetabular cup and augment is stable and primarily influenced by the size of the contact area, not the type and mechanical properties of bone cement.
This study provides a theoretical basis for the use of augments during revision surgery. The factors affecting the strength were explored. This could promote the improvement of prostheses and guide the design of a new generation of prostheses.
对于严重骨缺损的髋关节翻修手术,使用多孔金属髋臼组件并联合使用增强物是一种潜在的方法。然而,此前尚未对增强物 - 髋臼杯界面的强度进行研究。本研究评估了髋臼杯与增强物之间的界面强度。
采用撬出试验测试增强物、骨水泥和髋臼杯之间界面的强度。研究了接触面积和骨水泥类型对界面强度的影响。
界面表现出优异的稳定性。对于相同的髋臼杯 - 增强物组合以及不同性质的不同骨水泥,主要力学指标未见统计学差异(p > 0.05);然而,当使用相同骨水泥与不同的髋臼杯 - 增强物组合时,最大力矩和屈服力矩随着接触面积的减小而显著降低(p < 0.05)。界面刚度呈下降趋势,尽管差异无统计学意义。失效总是发生在骨水泥 - 髋臼杯一侧。
髋臼杯与增强物之间的界面稳定,主要受接触面积大小的影响,而非骨水泥的类型和力学性能。
本研究为翻修手术中使用增强物提供了理论依据。探讨了影响强度的因素。这有助于促进假体的改进并指导新一代假体的设计。