Factors that Enhance the Stability of Polidocanol and Monoethanolamine Oleic Acid Foams, and Identification of the Optimal Foam-Creating Device.

作者信息

Higashino Nobuyuki, Sonomura Tetsuo, Mimura Ryosuke, Fukuda Kodai, Sato Hirotatsu, Kawai Nobuyuki

机构信息

Department of Radiology, Wakayama Medical University, 811-1 Kimiidera, Wakayamashi, Wakayama, 641-8509, Japan.

出版信息

Cardiovasc Intervent Radiol. 2025 Aug;48(8):1174-1187. doi: 10.1007/s00270-025-04072-0. Epub 2025 Jun 9.

Abstract

PURPOSE

To investigate the effects of the air ratio and the addition of glucose solution (Gluc) on the stability of polidocanol-nonionic water-soluble contrast medium (Po/CM) and monoethanolamine oleate-nonionic water-soluble contrast medium (EO/CM) foams, and to identify the optimal device to prepare these foams.

MATERIALS AND METHODS

Four sclerosants were used: Po/CM (Po:CM = 1:1), Po/CM-Gluc (Po:CM:70% Gluc = 2:1:1), EO/CM (EO:CM = 1:1), and EO/CM-Gluc (EO:CM:70% Gluc = 2:1:1). The four sclerosants were mixed with air at ratios of 1:4, 1:7, and 1:10 using the Micro Magic connector (MM), hydrophobic emulsifying connector (Hbic), and hydrophilic emulsifying connector (Hlic). The times taken to prepare the foams using each device and the foam half-lives were measured.

RESULTS

The Po/CM (1:10) and Po/CM-Gluc (1:10) foams could not be prepared using any device. The EO/CM-Gluc (1:10), Po/CM (1:7), and Po/CM-Gluc (1:7) foams could not be prepared using the MM. The times taken to prepare the foams were not significantly different between the Hbic and Hlic for EO/CM-Gluc (1:4) and Po/CM (1:4). The times taken to prepare the other foams decreased significantly in the order MM > Hbic > Hlic. The half-lives of the foams were significantly prolonged by increasing the air ratio. For foams with the same air ratio, the presence of Gluc significantly increased the half-lives of EO/CM and Po/CM compared with the foams without Gluc.

CONCLUSION

The half-lives of foams were prolonged by increasing their air ratio and by adding Gluc. Of the tested mixture devices, Hlic created foams with longer half-lives in the shortest time.

摘要

目的

研究空气比例及葡萄糖溶液(Gluc)的添加对聚多卡醇 - 非离子型水溶性造影剂(Po/CM)泡沫和油酸单乙醇胺 - 非离子型水溶性造影剂(EO/CM)泡沫稳定性的影响,并确定制备这些泡沫的最佳装置。

材料与方法

使用四种硬化剂:Po/CM(Po:CM = 1:1)、Po/CM - Gluc(Po:CM:70% Gluc = 2:1:1)、EO/CM(EO:CM = 1:1)和EO/CM - Gluc(EO:CM:70% Gluc = 2:1:1)。使用微型魔术连接器(MM)、疏水乳化连接器(Hbic)和亲水乳化连接器(Hlic),将这四种硬化剂与空气按1:4、1:7和1:10的比例混合。测量使用每种装置制备泡沫所需的时间以及泡沫的半衰期。

结果

使用任何装置均无法制备Po/CM(1:10)和Po/CM - Gluc(1:10)泡沫。使用MM无法制备EO/CM - Gluc(1:10)、Po/CM(1:7)和Po/CM - Gluc(1:7)泡沫。对于EO/CM - Gluc(1:4)和Po/CM(1:4),Hbic和Hlic制备泡沫所需的时间无显著差异。制备其他泡沫所需的时间按MM>Hbic>Hlic的顺序显著减少。通过增加空气比例,泡沫的半衰期显著延长。对于具有相同空气比例的泡沫,与未添加Gluc的泡沫相比,Gluc的存在显著增加了EO/CM和Po/CM的半衰期。

结论

通过增加空气比例和添加Gluc可延长泡沫的半衰期。在所测试的混合装置中,Hlic在最短时间内制备出半衰期更长的泡沫。