Speckle-Tracking Echocardiography in Dogs: Evaluating Imaging Parameters and Methodological Variability in Global Longitudinal Strain Assessment.

作者信息

Mogensen Jonas E, Bach Maiken B T, Bay Pernille G, Varlik Tuğba, Willesen Jakob L, Gleerup Caroline H, Koch Jørgen

机构信息

Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Dyrlægevej 16, 1870 Frederiksberg, Denmark.

Department of Veterinary Medicine, Internal Medicine, Bursa Uludağ University, Görükle Kampüsü, Görükle, 16059 Nilüfer, Bursa, Türkiye.

出版信息

Animals (Basel). 2025 May 23;15(11):1523. doi: 10.3390/ani15111523.

Abstract

Two-dimensional speckle-tracking echocardiography (2D-STE) is an advanced imaging technique that offers quantitative insights into myocardial function by analyzing the motion of speckles created during ultrasound-tissue interactions. This study aims to evaluate the reliability of 2D-STE by examining the impact of key technical parameters on global longitudinal strain (GLS) measurement accuracy and comparing two speckle-tracking analysis methods provided by GE Healthcare: quantitative analysis of the 2D strain (2D strain) and automated function imaging (AFI). The prospective study consisted of two cohorts. In the first cohort, including 16 healthy dogs, the influence of frame rate, heart rate variation, zoom, transducer frequency, and image foreshortening on speckle-tracking values was assessed. In the second cohort, which included 10 healthy dogs, 2D-STE parameters were obtained with the 2D strain and AFI to assess agreement between the methods and observer variability. Our findings indicate that foreshortening ( < 0.01, Cohen's d: 0.52, CI: -17.81 to -24.83) and heart rate variability ( = 0.02, Cohen's d: 0.72, CI: -18.07 to -26.23) significantly affect speckle-tracking measurements. While zoom, frame rate, and frequency did not show a significant impact. Additionally, while the 2D strain and AFI exhibited a strong correlation, a significant systematic bias was identified, with AFI underestimating strain values compared to the 2D strain. Intra- and inter-observer coefficients of variation (CV) were below 9% for both methods, supporting their reliability. These findings emphasize the need to optimize image acquisition and selection criteria, which enhances the accuracy and reliability of the speckle-tracking analysis.

摘要

二维斑点追踪超声心动图(2D-STE)是一种先进的成像技术,通过分析超声与组织相互作用过程中产生的斑点运动,对心肌功能进行定量分析。本研究旨在通过检查关键技术参数对整体纵向应变(GLS)测量准确性的影响,并比较通用电气医疗集团提供的两种斑点追踪分析方法:二维应变定量分析(2D应变)和自动功能成像(AFI),来评估2D-STE的可靠性。这项前瞻性研究包括两个队列。在第一个队列中,有16只健康犬,评估了帧率、心率变化、缩放、换能器频率和图像缩短对斑点追踪值的影响。在第二个队列中,有10只健康犬,使用2D应变和AFI获得2D-STE参数,以评估两种方法之间的一致性和观察者变异性。我们的研究结果表明,图像缩短(<0.01,科恩d值:0.52,置信区间:-17.81至-24.83)和心率变异性(=0.02,科恩d值:0.72,置信区间:-18.07至-26.23)显著影响斑点追踪测量。而缩放、帧率和频率未显示出显著影响。此外,虽然2D应变和AFI表现出很强的相关性,但发现存在显著的系统偏差,与2D应变相比,AFI低估了应变值。两种方法的观察者内和观察者间变异系数(CV)均低于9%,支持了它们的可靠性。这些发现强调了优化图像采集和选择标准的必要性,这可以提高斑点追踪分析的准确性和可靠性。