Validation and evaluation of visual acuity and contrast sensitivity with a new digital device: OPTOTAB®.

作者信息

Ponce-García Víctor, Silva-Viguera María-Carmen, García-Romera Marta-C, Guerra-Sancho Yael, Heredia-Ríos Noelia, Bautista-Llamas María-José

机构信息

Department of Physics of Condensed Matter, Optics Area, Vision Research Group (CIVIUS), University of Seville, Seville, Spain.

出版信息

Ophthalmic Physiol Opt. 2025 Sep;45(6):1302-1316. doi: 10.1111/opo.13536. Epub 2025 Jun 13.

Abstract

PURPOSE

To validate a new digital device: Optotab®+ (SmarThings4Vision) for measuring visual acuity (VA) and contrast sensitivity (CS).

METHODS

A validation study was conducted involving 20 healthy subjects (aged: 18-29 years). Distance and near VA and CS were assessed using the Optotab®+ and compared with the ETDRS test and Sloan letters for distance VA, the ETDRS test and LEA numbers® near vision card for near VA and the Regan and CSV-1000 tests for CS.

RESULTS

No significant differences were noted between the right eye (RE) and left eye (L for distance (p > 0.99) and near VA (p = 0.32) or test-retest measurements (p > 0.05) using the Optotab®+. Distance VA differed significantly from the ETDRS (mean differences: -0.07 and -0.07 logMAR for the RE and LE, respectively, p = 0.01). Intraclass correlation coefficients (ICC) indicated moderate reliability (RE: 0.64, LE: 0.48), while the ANOVA showed no significant differences (p = 0.75 for RE and p = 0.58 for LE). For CS, no significant differences were found between the RE and LE, except at 12 cpd (p = 0.04). ICC was highest at 18 cpd (0.90). A significant test-retest difference was observed at 6 cpd for the RE (0.10 log units, p = 0.01). Comparisons showed significant differences at 6 cpd between Optotab®+ and the CSV-1000, and at 3, 12 and 18 cpd between Optotab®+ and the Regan test. ICC indicated low to high reliability across spatial frequencies, while the ANOVA did not show significant differences or low variability. No significant inter-examiner differences were identified in VA and CS (p > 0.05), confirming strong reproducibility.

CONCLUSIONS

This study demonstrates that the Optotab®+ has moderate reliability for VA assessment and higher precision at higher spatial frequencies for CS, suggesting its clinical utility. Outcomes regarding reliability, repeatability and reproducibility support its validity as an effective tool for measuring visual parameters in clinical practice.

摘要

目的

验证一种新型数字设备Optotab®+(SmarThings4Vision)用于测量视力(VA)和对比敏感度(CS)的效果。

方法

进行了一项验证研究,纳入20名健康受试者(年龄:18 - 29岁)。使用Optotab®+评估远视力和近视力以及CS,并与ETDRS测试、用于远视力的Sloan字母表、用于近视力的ETDRS测试和LEA数字近视力卡以及用于CS的Regan和CSV - 1000测试进行比较。

结果

使用Optotab®+时,右眼(RE)和左眼(LE)在远视力(p > 0.99)和近视力(p = 0.32)或重测测量(p > 0.05)方面未发现显著差异。远视力与ETDRS有显著差异(RE和LE的平均差异分别为 - 0.07和 - 0.07 logMAR,p = 0.01)。组内相关系数(ICC)表明可靠性中等(RE:0.64,LE:0.48),而方差分析未显示显著差异(RE为p = 0.75,LE为p = 0.58)。对于CS,除了在12 cpd时(p = 0.04),RE和LE之间未发现显著差异。ICC在18 cpd时最高(0.90)。在RE的6 cpd处观察到显著的重测差异(0.10 log单位,p = 0.01)。比较显示Optotab®+与CSV - 1000在6 cpd时有显著差异,Optotab®+与Regan测试在3、12和18 cpd时有显著差异。ICC表明在不同空间频率下可靠性从低到高,而方差分析未显示显著差异或低变异性。在视力和对比敏感度方面未发现显著的检查者间差异(p > 0.05),证实了很强的可重复性。

结论

本研究表明Optotab®+在视力评估方面具有中等可靠性,在较高空间频率下对比敏感度测量具有更高精度,表明其临床实用性。关于可靠性、重复性和可重复性的结果支持其作为临床实践中测量视觉参数的有效工具的有效性。