Suppr超能文献

使用RS-1光谱域光学相干断层扫描和OA-2000生物测量仪进行眼轴长度测量的比较。

Comparison of axial length measurements with RS-1 spectral domain optical coherence tomography and OA-2000 biometer.

作者信息

Fujihara Hitomi, Kakihara Shinji, Hirano Takao, Murata Toshinori

机构信息

Department of Ophthalmology, Shinshu University School of Medicine, Matsumoto, Nagano 390-8621, Japan.

出版信息

Int J Ophthalmol. 2025 Jun 18;18(6):1085-1089. doi: 10.18240/ijo.2025.06.15. eCollection 2025.

Abstract

AIM

To determine the utility of the RS-1 spectral domain optical coherence tomography (SD-OCT), which incorporates an "OCT Analysis Correction Parameter" for approximating axial length, in comparison with measurements obtained from the OA-2000.

METHODS

Twenty-five right eyes of healthy individuals were included. Two horizontal line scans were conducted using the RS-1, and OCT Analysis Correction Parameters were recorded. Axial length was measured twice per eye using the OA-2000. Correlation between devices was assessed using Pearson correlation coefficient, coefficient of repeatability (CR%), and Bland-Altman analysis.

RESULTS

High correlation was found between axial length measurements from RS-1 and OA-2000 (=0.986, <0.0001). The coefficient of repeatability was 1.56% for RS-1 and 0.115% for OA-2000. Mean axial length was 25.12±1.38 mm for RS-1 and 24.90±1.54 mm for OA-2000, with RS-1 showing a statistically significant larger value (paired -test, =0.0009).

CONCLUSION

The axial length measurements of RS-1 demonstrate a strong positive correlation with those from the OA-2000. This indicates that it could potentially be used to forecast scan ranges and choose suitable databases according to axial length categories.

摘要

目的

与OA - 2000所测数据相比,确定采用“OCT分析校正参数”来估算眼轴长度的RS - 1谱域光学相干断层扫描(SD - OCT)的效用。

方法

纳入25名健康个体的右眼。使用RS - 1进行两次水平线扫描,并记录OCT分析校正参数。使用OA - 2000对每只眼睛的眼轴长度测量两次。采用Pearson相关系数、重复性系数(CR%)和Bland - Altman分析评估设备之间的相关性。

结果

RS - 1和OA - 2000所测眼轴长度之间存在高度相关性(=0.986,<0.0001)。RS - 1的重复性系数为1.56%,OA - 2000的重复性系数为0.115%。RS - 1测得的平均眼轴长度为25.12±1.38mm,OA - 2000测得的平均眼轴长度为24.90±1.54mm,RS - 1显示出具有统计学意义的更大值(配对检验,=0.0009)。

结论

RS - 1的眼轴长度测量结果与OA - 2000的测量结果显示出强正相关。这表明它有可能用于预测扫描范围,并根据眼轴长度类别选择合适的数据库。

相似文献

4
Prognostic factors for return to work in breast cancer survivors.乳腺癌幸存者恢复工作的预后因素。
Cochrane Database Syst Rev. 2025 May 7;5(5):CD015124. doi: 10.1002/14651858.CD015124.pub2.
5
Non-pharmacological interventions for sleep promotion in hospitalized children.促进住院儿童睡眠的非药物干预措施。
Cochrane Database Syst Rev. 2022 Jun 15;6(6):CD012908. doi: 10.1002/14651858.CD012908.pub2.
6
Mucolytics for children with chronic suppurative lung disease.用于患有慢性化脓性肺病儿童的黏液溶解剂。
Cochrane Database Syst Rev. 2025 Mar 28;3(3):CD015313. doi: 10.1002/14651858.CD015313.pub2.
9
Electronic cigarettes for smoking cessation.用于戒烟的电子烟。
Cochrane Database Syst Rev. 2025 Jan 29;1(1):CD010216. doi: 10.1002/14651858.CD010216.pub9.

本文引用的文献

1
Photoreceptor assessment in age-related macular degeneration.年龄相关性黄斑变性中的光感受器评估
Eye (Lond). 2025 Feb;39(2):284-295. doi: 10.1038/s41433-024-03462-x. Epub 2024 Nov 22.
3
Optical coherence tomography angiography in diabetic retinopathy.光学相干断层扫描血管造影在糖尿病视网膜病变中的应用。
Prog Retin Eye Res. 2023 Nov;97:101206. doi: 10.1016/j.preteyeres.2023.101206. Epub 2023 Jul 26.
7
10
Optical Coherence Tomography Findings in Infectious Posterior Uveitis.感染性后部葡萄膜炎的光学相干断层扫描表现。
Ocul Immunol Inflamm. 2022 Apr 3;30(3):652-663. doi: 10.1080/09273948.2022.2032197. Epub 2022 Feb 28.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验