Fujihara Hitomi, Kakihara Shinji, Hirano Takao, Murata Toshinori
Department of Ophthalmology, Shinshu University School of Medicine, Matsumoto, Nagano 390-8621, Japan.
Int J Ophthalmol. 2025 Jun 18;18(6):1085-1089. doi: 10.18240/ijo.2025.06.15. eCollection 2025.
To determine the utility of the RS-1 spectral domain optical coherence tomography (SD-OCT), which incorporates an "OCT Analysis Correction Parameter" for approximating axial length, in comparison with measurements obtained from the OA-2000.
Twenty-five right eyes of healthy individuals were included. Two horizontal line scans were conducted using the RS-1, and OCT Analysis Correction Parameters were recorded. Axial length was measured twice per eye using the OA-2000. Correlation between devices was assessed using Pearson correlation coefficient, coefficient of repeatability (CR%), and Bland-Altman analysis.
High correlation was found between axial length measurements from RS-1 and OA-2000 (=0.986, <0.0001). The coefficient of repeatability was 1.56% for RS-1 and 0.115% for OA-2000. Mean axial length was 25.12±1.38 mm for RS-1 and 24.90±1.54 mm for OA-2000, with RS-1 showing a statistically significant larger value (paired -test, =0.0009).
The axial length measurements of RS-1 demonstrate a strong positive correlation with those from the OA-2000. This indicates that it could potentially be used to forecast scan ranges and choose suitable databases according to axial length categories.
与OA - 2000所测数据相比,确定采用“OCT分析校正参数”来估算眼轴长度的RS - 1谱域光学相干断层扫描(SD - OCT)的效用。
纳入25名健康个体的右眼。使用RS - 1进行两次水平线扫描,并记录OCT分析校正参数。使用OA - 2000对每只眼睛的眼轴长度测量两次。采用Pearson相关系数、重复性系数(CR%)和Bland - Altman分析评估设备之间的相关性。
RS - 1和OA - 2000所测眼轴长度之间存在高度相关性(=0.986,<0.0001)。RS - 1的重复性系数为1.56%,OA - 2000的重复性系数为0.115%。RS - 1测得的平均眼轴长度为25.12±1.38mm,OA - 2000测得的平均眼轴长度为24.90±1.54mm,RS - 1显示出具有统计学意义的更大值(配对检验,=0.0009)。
RS - 1的眼轴长度测量结果与OA - 2000的测量结果显示出强正相关。这表明它有可能用于预测扫描范围,并根据眼轴长度类别选择合适的数据库。