Quantifying performance enhancements in photon-counting pediatric head CT: a direct comparison with energy-integrating CT.

作者信息

Naceanceno Kevin, Thomas Allan, Zhang Zhongwei, Sanchez Adrian, Lappe Christi, Skinner Nathan, Abadia Andres, Landrum Kathryn, McKinstry Robert

机构信息

Department of Radiology, Washington University in St. Louis, Mallinckrodt (United States), Campus Box 8131-19-01, 510 South Kingshighway Boulevard, Hazelwood, St. Louis, MO, 63110, United States.

St. Louis Children's Hospital, St Louis, United States.

出版信息

Pediatr Radiol. 2025 Jun 20. doi: 10.1007/s00247-025-06295-2.

Abstract

BACKGROUND

Given the unique detector design, photon-counting detector CT (PCD-CT) systems may reduce dose and improve image quality in pediatric head CT.

OBJECTIVE

This retrospective study compared image noise characteristics, spatial resolution, and radiation dose between PCD-CT and energy-integrating detector CT (EID-CT) in pediatric non-contrast head CT exams. Clinical data was compared with results from phantom studies.

MATERIALS AND METHODS

Radiation dose, image noise, and signal-to-noise ratio (SNR) differences were compared in matched clinical EID-CT vs. PCD-CT pairs from 162 full-dose and 33 low-dose pediatric head CT exams. Image noise, spatial resolution, and contrast-to-noise characteristics were also compared between EID-CT and PCD-CT using preclinical phantoms at different slice thicknesses. The ability of phantom scans to predict clinically observed performance changes between EID-CT and PCD-CT was also analyzed.

RESULTS

In full-dose pediatric head CT studies, PCD-CT demonstrated 24% lower noise and 18% higher SNR at equal slice thickness (both P < 0.001), as well as 2.3% lower CTDI (P = 0.03). In low-dose cases, PCD-CT showed 20% lower noise, 13% higher SNR, and 12.5% lower CTDI (all P < 0.001). Phantom studies closely agreed with clinical data and showed PCD-CT had ~ 19% lower image noise than EID-CT at the same dose and slice thickness using clinical scan and reconstruction parameters.

CONCLUSION

Relative to EID-CT, PCD-CT offers similar noise texture with reduced noise magnitude at comparable or reduced radiation doses and thinner slices in pediatric head CT. When combined with improved spatial resolution, all these results point to the ability of PCD-CT to enhance imaging performance in pediatric head CT.

摘要

背景

鉴于独特的探测器设计,光子计数探测器CT(PCD-CT)系统可能会降低儿科头部CT的辐射剂量并提高图像质量。

目的

这项回顾性研究比较了儿科非增强头部CT检查中PCD-CT与能量积分探测器CT(EID-CT)之间的图像噪声特征、空间分辨率和辐射剂量。将临床数据与体模研究结果进行了比较。

材料与方法

在162例全剂量和33例低剂量儿科头部CT检查中,对匹配的临床EID-CT与PCD-CT对的辐射剂量、图像噪声和信噪比(SNR)差异进行了比较。还使用不同层厚的临床前体模比较了EID-CT和PCD-CT之间的图像噪声、空间分辨率和对比噪声特征。还分析了体模扫描预测EID-CT和PCD-CT之间临床观察到的性能变化的能力。

结果

在全剂量儿科头部CT研究中,在相同层厚下,PCD-CT的噪声降低了24%,SNR提高了18%(均P <0.001),CTDI降低了2.3%(P =0.03)。在低剂量情况下,PCD-CT的噪声降低了20%,SNR提高了13%,CTDI降低了12.5%(均P <0.001)。体模研究与临床数据密切一致,表明在相同剂量和层厚下,使用临床扫描和重建参数时,PCD-CT的图像噪声比EID-CT低约19%。

结论

相对于EID-CT,PCD-CT在儿科头部CT中以可比或更低的辐射剂量和更薄的层厚提供了类似的噪声纹理,同时降低了噪声幅度。结合提高的空间分辨率,所有这些结果表明PCD-CT能够增强儿科头部CT的成像性能。