Warner G T, Potter C G, Yrjonen T, Soini E
Int J Appl Radiat Isot. 1985 Oct;36(10):819-21. doi: 10.1016/0020-708x(85)90034-1.
A new design for a liquid scintillation counter based on a flat-bed geometry is described. Micro-samples are dried or filtered onto transfer membranes or glass fibre filters in a 6 X 16 matrix, compatible with 96-well micro-titration plate filtration assays of labelled cells. A prototype counter without lead shielding had low background countrates (2-3 cpm for 3H) giving a figure of merit of 1325 (and 1292 for 14C). Only 5-15 ml of scintillant/96 samples are required and thus the volume of radioactive waste is low.