Suppr超能文献

爱丽丝NightOne单带式监测仪对阻塞性睡眠呼吸暂停的诊断准确性及无线数据传输的可靠性

Diagnostic Accuracy of the Alice NightOne Single-Belt Monitor for Obstructive Sleep Apnea and Reliability of Wireless Data Transfer.

作者信息

Peng MaoHuan, Zhang YuanYuan, Zhao Rui, Li Jing, Zhang XueLi, Deng LiHua, Zhao Long, Zhou Bing, Wang XinRu, Dong XiaoSong, Han Fang

机构信息

Department of Sleep Medicine, Peking University People's Hospital, Beijing, People's Republic of China.

Department of General Practice, Peking University People's Hospital, Beijing, People's Republic of China.

出版信息

Nat Sci Sleep. 2025 Jun 26;17:1493-1505. doi: 10.2147/NSS.S527805. eCollection 2025.

Abstract

OBJECTIVE

To validate the performance of a single-belt type 3 portable monitor Alice NightOne for diagnosis of obstructive sleep apnea and its reliability in remote data transmission.

METHODS

Our study included two parts: in-laboratory Alice NightOne (ANO) monitoring (ANO) and home sleep apnea testing (HSAT) using ANO (ANO). For ANO, the participants underwent polysomnography (PSG) with ANO simultaneously. For ANO, the participants completed unattended overnight ANO out of sleep center and PSG was performed in another night. The ANO recordings were transmitted to cloud database wirelessly in addition to traditional wire transmission. Message digest-5 (MD5) algorithm was utilized to verify the integrity of the cloud data.

RESULTS

Ninety-one ANO and 170 ANO recordings were analyzed. Both the respiratory event index (REI) on ANO and that on ANO were lower than the corresponding apnea-hypopnea index (AHI) on PSG (24.9 ± 20.5 events/h vs 31.6 ± 25.0 events/h, and 26.7 ± 17.0 events/h vs 35.3 ± 21.2 events/h respectively, P < 0.001). Bland-Altman analysis of REI on ANO versus AHI on PSG showed a mean difference (95% confidence interval) of -6.7 (-8.4, -4.9) events/h. For REI on ANO versus AHI on PSG, the difference is -8.0 (-9.9, -6.0) events/h. With threshold of REI ≥ 5 events/h for OSA diagnosis, ANO had 98.8% sensitivity, 90.0% specificity, 99.4% positive predictive value. The MD5 algorithm verified the identity between uploaded cloud data and original data.

CONCLUSION

With single thoracoabdominal belt, Alice NightOne can help diagnosis of obstructive sleep apnea with good sensitivity and specificity, though it may underestimate AHI. Furthermore, it provides reliable support based on solid data teletransmission and scoring synchronization, which may increase the ability of diagnosis and management of OSA through telemedicine.

摘要

目的

验证单带式3型便携式监测仪Alice NightOne用于诊断阻塞性睡眠呼吸暂停的性能及其在远程数据传输中的可靠性。

方法

我们的研究包括两个部分:实验室Alice NightOne(ANO)监测(ANO)和使用ANO进行的家庭睡眠呼吸暂停测试(HSAT)(ANO)。对于ANO,参与者同时接受ANO多导睡眠图(PSG)检查。对于ANO,参与者在睡眠中心外完成无人值守的夜间ANO检查,PSG在另一个晚上进行。除了传统的有线传输外,ANO记录还被无线传输到云数据库。使用消息摘要-5(MD5)算法验证云数据的完整性。

结果

分析了91份ANO记录和170份ANO记录。ANO上的呼吸事件指数(REI)和ANO上的呼吸事件指数均低于PSG上相应的呼吸暂停低通气指数(AHI)(分别为24.9±20.5次/小时对31.6±25.0次/小时,以及26.7±17.0次/小时对35.3±21.2次/小时,P<0.001)。ANO上的REI与PSG上的AHI的Bland-Altman分析显示平均差异(95%置信区间)为-6.7(-8.4,-4.9)次/小时。对于ANO上的REI与PSG上的AHI,差异为-8.0(-9.9,-6.0)次/小时。以REI≥5次/小时作为阻塞性睡眠呼吸暂停(OSA)诊断阈值,ANO的灵敏度为98.8%,特异度为90.0%,阳性预测值为99.4%。MD5算法验证了上传的云数据与原始数据的一致性。

结论

采用单胸腹带,Alice NightOne虽可能低估AHI,但能以良好的灵敏度和特异度辅助诊断阻塞性睡眠呼吸暂停。此外,它基于可靠的数据远程传输和评分同步提供可靠支持,这可能增强通过远程医疗对OSA的诊断和管理能力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/7688/12209879/c43834eb79e7/NSS-17-1493-g0001.jpg