Mesgarzadeh M, Revesz G, Bonakdarpour A, Betz R R
Skeletal Radiol. 1985;14(3):205-6. doi: 10.1007/BF00355565.
Forces and torques, due to a 0.3 Tesla magnetic field were evaluated on ten hip prostheses and ten hemostat clips. Measurements were performed with an instrument utilizing the movement of a laser beam caused by the deflection of a cantilever. The results indicate effects, if any, to be smaller than the instrument's sensitivity which, at its highest, was 7 mg of force and 125 mg.cm of torque.
对十个髋关节假体和十个止血夹,评估了0.3特斯拉磁场产生的力和扭矩。测量是使用一种仪器进行的,该仪器利用悬臂梁偏转引起的激光束移动。结果表明,即使存在影响,其影响也小于仪器的灵敏度,该仪器的最高灵敏度为7毫克力和125毫克·厘米扭矩。